Z-2237-2010

NeuViz Dual series Computed Tomography (CT) Scanner System, X-Ray System. Part Number: 989605651321. Intended...

Neuisys, LLC / initiated 2008-07-21 / Terminated / root cause: software

Neuisys, LLC began a recall of NeuViz Dual series Computed Tomography (CT) Scanner System, X-Ray System. Part Number: 989605651321. Intended to produce cross-section images of head and body... on . The reason FDA records is "The reference lines for image generated for a Surview scan may appear in the incorrect position in the "FILM" display mode". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-2237-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Neuisys, LLC
Product
NeuViz Dual series Computed Tomography (CT) Scanner System, X-Ray System. Part Number: 989605651321. Intended to produce cross-section images of head and body by computer reconstruction of X-ray transmission data taken at different angles.
Product code
JAK System, X-Ray, Tomography, Computed
Reason for recall
The reference lines for image generated for a Surview scan may appear in the incorrect position in the "FILM" display mode.
Root cause tag
software
FDA root cause
Software design
Action
On 07/21/2008, NEUISYS Imaging Systems Solutions distributor in Greensboro, NC began sending the URGENT DEVICE CORRECTION letter on behalf of Philips and NeuSoft Medical Systems (PNMS), Co., Ltd in China. The letter informs the consignees that there is potential safety problem in the NeuViz Dual series CT scanners due to the malfunction of "Incorrect Reference Line in the Film". When a "Head first, Couch out" or a "Feet first, Couch in" Surview scan is started then paused midstream, the image generated is what the doctor expects, but the reference lines appear in the incorrect position in "FILM" display mode. The acquired images accurately represent the exact positions planned on the original view. The incorrect/unmatched reference line in film may result in incorrect diagnosis. The consignees are advised not to stop or pause during Surview acquisition while PNMS in China will provide the updated software patch to all affected systems free of charge. The Field Change Order #FCOP-08-CT-NMS-008 SP4 will be implemented by NEUISYS Imaging Systems Solutions in Greensboro, NC., who is a distributor for PNMS. Consignees are advised to e-mail Service Support Department of PNMS at [email protected] regarding the recall.
Quantity in commerce
19 units in the US
Distribution
Nationwide Distribution
Product codes and lots
Serial Number:  400562, 400563, 400626, NDH009EI, NDH011El, NDH012EI, NDH013EI, NDH016EI, NDH017EI, NDH025EI, NDH026EI, NDH027EI, NDH028EI, NDH030EI, NDH031EI, NDH034EI, NDHR080001, NDHR08002, and NDHRO80003.