BD 60 ml Syringe, Luer-Lok Tip Sterile; Cases : 160 (4 x 40) Shelf Carton: 40 BD; Franklin Lakes, NJ 07417...
Becton Dickinson & Co / initiated 2008-07-15 / Terminated / root cause: packaging
Becton Dickinson & Company began a recall of BD 60 ml Syringe, Luer-Lok Tip Sterile; Cases : 160 (4 x 40) Shelf Carton: 40 BD; Franklin Lakes, NJ 07417 USA. Made in USA. Syringes for use in aspiration and... on . The reason FDA records is "Becton Dickinson identified the presence of open seals during an internal inspection process". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-2374-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Becton Dickinson & Co
- Product
- BD 60 ml Syringe, Luer-Lok Tip Sterile; Cases : 160 (4 x 40) Shelf Carton: 40 BD; Franklin Lakes, NJ 07417 USA. Made in USA. Syringes for use in aspiration and injection of medications.
- Product code
- FMF Syringe, Piston
- Reason for recall
- Becton Dickinson identified the presence of open seals during an internal inspection process.
- Root cause tag
- packaging
- FDA root cause
- Package design/selection
- Action
- Urgent: Product Recall letters were sent to Distributors and Syringe Customers on July 15, 2008 by UPS second day air. Customers are to immediately review their inventory, determine if they have any affected product, and remove the affected if they have any. Next customers should complete the enclosed Customer Packing List and BD will send a replacement product upon receipt of the returned product. Questions are to be directed to BD Customer Service at 1-888-237-2762, M-F, 8:00 AM to 5:00 PM EST.
- Quantity in commerce
- 1246 lots
- Distribution
- Nationwide Distribution.
- Product codes and lots
- Reorder Number: 309653; Lot numbers beginning with the numbers 5, 6, or 7 and also the following lot numbers. 8010954, 8014571, 8031115, 8031117, 8031119, 8031124, 8031133, 8038154, 8038158, 8045564, 8045567, 8058890, 8058896, 8058898, 8058900, 8058902, 8072401, 8072409, 8072414, 8072419, 8079645, and 8079656. Note: In order to make the recall practical for customers to implement, BD has requested the return of lots beginning with the numbers 5,6 or 7. while not all lots beginning with the digits 5,6 and 7 exhibit the issue, it is much easier to instruct customers to return these lots, rather than have customrs examine/check a more comprehensive list of affected and unaffected lots.