Z-1197-2009

Beckman Coulter Access Immunoassay System Hybritech PSA Calibrators, Product Number: 37205. The Access...

Siemens Healthcare Diagnostics Inc. / initiated 2008-01-08 / Terminated / root cause: packaging

Beckman Coulter Inc began a recall of Beckman Coulter Access Immunoassay System Hybritech PSA Calibrators, Product Number: 37205. The Access Hybritech PSA assay is a paramagnetic particle... on . The reason FDA records is "Beckman Coulter incorrectly shipped Access Hybritech PSA calibrators containing dual calibration cards (Hybritech and WHO) to some Access Hybritech PSA Customers. The WHO calibration for PSA is not currently approved in...". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-1197-2009
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
Beckman Coulter Access Immunoassay System Hybritech PSA Calibrators, Product Number: 37205. The Access Hybritech PSA assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of total prostate specific antigen (PSA) levels in human serum using the Access Immunoassay Systems.
Product code
LTJ Prostate-Specific Antigen (Psa) For Management Of Prostate Cancers
Reason for recall
Beckman Coulter incorrectly shipped Access Hybritech PSA calibrators containing dual calibration cards (Hybritech and WHO) to some Access Hybritech PSA Customers. The WHO calibration for PSA is not currently approved in the United States.
Root cause tag
packaging
FDA root cause
Packaging process control
Action
An "Urgent: Product Corrective Action" letter dated January 10, 2008 was issued via fax or mail to customers. The letter described the issue and instructed consignees to discontinue use and discard the affected product. Consignees were also asked to complete and return a "Fax-Back Form" via fax (1-952-368-7610) or mail to Beckman Coulter, Inc. within 10 days of receiving the notification letter. Further questions should be addressed to Beckman Coulter Customer Technical Support at 1-800-854-3633.
Quantity in commerce
123 Units
Distribution
Nationwide Distribution -- Including states of AL, AZ, CA, FL, GA, IA, IL, IN, KS, KY, MA, MD, MI, MN, MO, NC, NY, OK, OR, PA, SC, TN, TX, VA, WI and WV.
Product codes and lots
Lot Numbers: 715292 and 722462.