Z-1344-2013Class III

Cardinal Health hCG Cassette Rapid Test, Part Number: B1077-22 (Mfg PN: 4582105026) The SP hCG Cassette Rapid...

Alere San Diego, Inc. / initiated 2013-01-21 / Terminated / root cause: labeling and UDI

Alere San Diego, Inc. began a recall of Cardinal Health hCG Cassette Rapid Test, Part Number: B1077-22 (Mfg PN: 4582105026) The SP hCG Cassette Rapid Test is a rapid chromatographic immunoassay for... on . FDA classified it as Class III and gives the reason as "Alere San Diego Inc. is recalling the Cardinal Health hCG Cassette Rapid Test because some of them were incorrectly packaged into Cardinal Health hCG Combo Rapid test kit boxes. The kit box incorrectly indicates the...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-1344-2013
Classification
Class III
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Alere San Diego, Inc.
Product
Cardinal Health hCG Cassette Rapid Test, Part Number: B1077-22 (Mfg PN: 4582105026) The SP hCG Cassette Rapid Test is a rapid chromatographic immunoassay for the qualitative detection of human chorionic gonadotropin (hCG) in urine, as an aid in the early detection of pregnancy. This test is for professional use.
Product code
JHI Visual, Pregnancy Hcg, Prescription Use
Reason for recall
Alere San Diego Inc. is recalling the Cardinal Health hCG Cassette Rapid Test because some of them were incorrectly packaged into Cardinal Health hCG Combo Rapid test kit boxes. The kit box incorrectly indicates the product may be used for both urine and serum sample types.
Root cause tag
labeling and UDI
FDA root cause
Labeling mix-ups
Action
The firm, Alere, sent an "URGENT MEDICAL DEVICE RECALL" letter dated January 21, 2013 to all their customers. The letter describes the product, problems identified and the actions to be taken. Customers are instructed to discontinue use and/or sale of the product; immediately discard all units of the the product in accordance with your local regulations; share this information with your laboratory staff and retain this letter for your records; if you have forwarded the product to another laboratory, provide a copy of this letter to them and include the name, address and a contact for the laboratory on the enclosed Customer Verification Form; complete and return the Customer Verification Form within 10 days via fax to: Technical Service at 858-805-8457 or e-mail to [email protected]. Customers with any questions are instructed to contact Alere Technical Services at (866) 216-0094 or by email at [email protected].
Quantity in commerce
2,167 kits
Distribution
Nationwide distribution: USA including states of: AL, AK, AR, AZ, CA, CO, CT, DE, HI, IA, ID, IL, IN, FL, GA, KY, KS, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, UT, TN, TX, VA, WA, WI, WV AND WY.
Product codes and lots
Lot: hCG2080330

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1344-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1344-2013Class IIITerminatedVoluntary: Firm initiated