Apogee Intermittent Catheter, 8 Fr 16", Catalog/Ref No. 1018, Sterile, Latex free. The device is intended to...
Apogee Medical, Inc began a recall of Apogee Intermittent Catheter, 8 Fr 16", Catalog/Ref No. 1018, Sterile, Latex free. The device is intended to be used as a urinary incontinence device designed... on . The reason FDA records is "Due to a package seal defect the catheter tip may have been inadvertently cut which could cause trauma to the urethra upon use. Also the sterility of the product could be compromised". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-2391-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Robling Medical, LLC
- Product
- Apogee Intermittent Catheter, 8 Fr 16", Catalog/Ref No. 1018, Sterile, Latex free. The device is intended to be used as a urinary incontinence device designed to drain urine from the bladder.
- Product code
- GBM Catheter, Urethral
- Reason for recall
- Due to a package seal defect the catheter tip may have been inadvertently cut which could cause trauma to the urethra upon use. Also the sterility of the product could be compromised.
- Root cause tag
- packaging
- FDA root cause
- Packaging process control
- Action
- Distributors and Retailers that have received any of the Catalog Numbers listed in the letter titled URGENT MEDICAL DEVICE RECALL were notified on/about July 17, 2008. Recipients of the letter are being requested to review their inventory and return any of the applicable Catalog Items from the specific lot numbers listed in the recall letter. Distributors are being instructed to notify their customers and inform them of the recall with a copy of the recall letter. Their notification should instruct their customers to return the recalled product to them so that they could return the product to Apogee Medical LLC. In addition to the initial letter, follow-up letters, telephone calls and emails will be utilized to assess the effectiveness of the recall. A spreadsheet has been set up and will be maintained in order to easily identify the customers that have been contacted, those who have responded and those who have returned product. If you have questions, contact Diane Peper at 919-435-5409.
- Quantity in commerce
- 5400 pcs
- Distribution
- U.S. Nationwide and Canada
- Product codes and lots
- Lot number 7206