Z-2400-2008
PLAC Test ELISA kit, catalog number 90106, enzyme immunoassay for the quantitative determination of Lp-PLA2...
Diadexus, Inc. / initiated 2008-07-18 / Terminated
DiaDexus, Inc began a recall of PLAC Test ELISA kit, catalog number 90106, enzyme immunoassay for the quantitative determination of Lp-PLA2 in human plasma and serum, kit containing coated... on . The reason FDA records is "Product may give Lp-PLA2 values at up to 22% lower than accurate values". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-2400-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Diadexus, Inc.
- Product
- PLAC Test ELISA kit, catalog number 90106, enzyme immunoassay for the quantitative determination of Lp-PLA2 in human plasma and serum, kit containing coated microwell stripplate and reagents (12 strips, 1 set calibrators 1 to 6, 0.25 ml each, 20x wash buffer, 50 mL, conjugate 23 mL, TMB, 11 mL, Stop solution, 11 mL
- Product code
- NOE Test, System, Immunoassay, Lipoprotein-Associated Phospholipase A2
- Reason for recall
- Product may give Lp-PLA2 values at up to 22% lower than accurate values.
- FDA root cause
- Finished device change control
- Action
- All customers received communication of product correction dated July 18. 2008, via Federal Express, which included a cover letter, technical bulletin, specimen handling best practices cards, and revised package inserts. E-mail communication was also sent to all but seven customers. Contact diaDexus Technical Support at 1-877-752-2837 for assistance.
- Quantity in commerce
- 65 (total)
- Distribution
- US, OUS: UK and Germany.
- Product codes and lots
- All lots