Z-2400-2008

PLAC Test ELISA kit, catalog number 90106, enzyme immunoassay for the quantitative determination of Lp-PLA2...

Diadexus, Inc. / initiated 2008-07-18 / Terminated

DiaDexus, Inc began a recall of PLAC Test ELISA kit, catalog number 90106, enzyme immunoassay for the quantitative determination of Lp-PLA2 in human plasma and serum, kit containing coated... on . The reason FDA records is "Product may give Lp-PLA2 values at up to 22% lower than accurate values". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-2400-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Diadexus, Inc.
Product
PLAC Test ELISA kit, catalog number 90106, enzyme immunoassay for the quantitative determination of Lp-PLA2 in human plasma and serum, kit containing coated microwell stripplate and reagents (12 strips, 1 set calibrators 1 to 6, 0.25 ml each, 20x wash buffer, 50 mL, conjugate 23 mL, TMB, 11 mL, Stop solution, 11 mL
Product code
NOE Test, System, Immunoassay, Lipoprotein-Associated Phospholipase A2
Reason for recall
Product may give Lp-PLA2 values at up to 22% lower than accurate values.
FDA root cause
Finished device change control
Action
All customers received communication of product correction dated July 18. 2008, via Federal Express, which included a cover letter, technical bulletin, specimen handling best practices cards, and revised package inserts. E-mail communication was also sent to all but seven customers. Contact diaDexus Technical Support at 1-877-752-2837 for assistance.
Quantity in commerce
65 (total)
Distribution
US, OUS: UK and Germany.
Product codes and lots
All lots