Thin-Flex Venous Return Cannulae, 33/43 FR. x 39 cm, REF TF3343O, Sterile EO, Edwards Lifesciences LLC...
Edwards Lifesciences Research Medical, Inc. began a recall of Thin-Flex Venous Return Cannulae, 33/43 FR. x 39 cm, REF TF3343O, Sterile EO, Edwards Lifesciences LLC, Irvine, CA 92614 Intended for cannula drainage of the... on . The reason FDA records is "Reinforcement spring may detach from the cannula body resulting in reduced venous blood return flow". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2401-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Edwards Lifesciences LLC
- Product
- Thin-Flex Venous Return Cannulae, 33/43 FR. x 39 cm, REF TF3343O, Sterile EO, Edwards Lifesciences LLC, Irvine, CA 92614 Intended for cannula drainage of the superior and inferior vena cava during extracorporeal circulation for a duration of less than and equal to 6 hours.
- Product code
- DWF Catheter, Cannula And Tubing, Vascular, Cardiopulmonary Bypass
- Reason for recall
- Reinforcement spring may detach from the cannula body resulting in reduced venous blood return flow.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- All domestic consignees were notified by a Product Recall letter on 07/15/2008 via FedEx. They were instructed to check their inventory for any affected product and return all unused product from their stock to Edwards. Customers are to contact Customer Service Organization to obtain a Returned Goods Authorization number and replacement product. Once inventory has been verified, the attached form needs to be completed and faxed to Edwards so that customer awareness can be confirmed. International accounts were notified on 07/14/2008.
- Quantity in commerce
- 707 units
- Distribution
- Worldwide Distribution --- USA including states of AL, CA, LA, MI, MN, MO, MS, NC, ND, SC, TX, and WA, and countries of Brazil, Israel, Japan, and Netherlands.
- Product codes and lots
- Lot Number 58449974
FIRST ARTICLE / WEEKLY