Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2405-2008

Synchron Acetaminophen (ACTM) Reagent, Part Number 472169. ACTM Reagent when used in conjunction with...

Siemens Healthcare Diagnostics Inc. / initiated 2008-07-10 / Terminated / root cause: manufacturing process

Beckman Coulter Inc began a recall of Synchron Acetaminophen (ACTM) Reagent, Part Number 472169. ACTM Reagent when used in conjunction with SYNCHRON LX System(s), UniCel DxC 600/800 Systems and... on . The reason FDA records is "This recall was initiated after Beckman Coulter verified through testing that Acetaminophen reagents (lots #M702524 & M703276) manufactured with contaminated heparin has shown a negative bias in performance between the...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2405-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
Synchron Acetaminophen (ACTM) Reagent, Part Number 472169. ACTM Reagent when used in conjunction with SYNCHRON LX System(s), UniCel DxC 600/800 Systems and SYNCHRON Systems Drug Calibrator 2 set, is intended for quantitative determination of Acetaminophen concentration in human serum or plasma.
Product code
LDP Colorimetry, Acetaminophen
Reason for recall
This recall was initiated after Beckman Coulter verified through testing that Acetaminophen reagents (lots #M702524 & M703276) manufactured with contaminated heparin has shown a negative bias in performance between the assay made with contaminated and uncontaminated heparin.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
The initial customer notifications with fax back response forms were mailed via US Postal Service beginning on 07/21/2008. The letter asked their customers to please discontinue use and discard appropriately the affected lots of Synchron Acetaminophen Reagent (M703276 & M702524). Customers were also instructed to contact their Customer Service Representative at (800) 526-3821 in the United States or (800) 463-7828 in Canada or, if they are outside the US or Canada, please contact your local Beckman Coulter Representative to discuss replacement options. They were also instructed to complete and return the enclosed response form within 10 days. In addition, the customers were informed that the contaminated Heparin was used in the following Beckman Coulter, Inc., products: Synchron Theophylline (THE) Reagent (lot # M701460), Synchron Phenytoin (PHY) Reagent and Synchron C-Reactive Protein (CRP) Reagent (lot# M703222, M703266, M703267 & M703268), however Beckman Coulter has verified through testing that the performance of these products are not affected.
Quantity in commerce
7764 test kits
Distribution
Nationwide and Canada
Product codes and lots
Lot Numbers M703276 and M702524