Z-2414-2008
Burette Set w/150 ml burette w/15 um filter & inj site, inj site, with automatic shutoff 87 in. Item/Catalog...
B Braun Medical Inc / initiated 2008-06-30 / Terminated / root cause: packaging
B. Braun Medical, Inc. began a recall of Burette Set w/150 ml burette w/15 um filter & inj site, inj site, with automatic shutoff 87 in. Item/Catalog number 375113. The product is shipped 20 units per... on . The reason FDA records is "incorrect burette was packaged with the product". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-2414-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- B Braun Medical Inc
- Product
- Burette Set w/150 ml burette w/15 um filter & inj site, inj site, with automatic shutoff 87 in. Item/Catalog number 375113. The product is shipped 20 units per carton. The product is for use after activation of shutoff valve (re-floating of disk)
- Product code
- FPA Set, Administration, Intravascular
- Reason for recall
- incorrect burette was packaged with the product
- Root cause tag
- packaging
- FDA root cause
- Mixed-up of materials/components
- Action
- The firm issued a Product Removal Notice dated 6/20/08 via USPS Certified Mail with registered receipt mail to all customers in receipt of affected product. The notice informs customers of the problem, to cease use and distribution immediately, and to return the product. Contact B. Braun Medical at 1-800-227-2862 for assistance.
- Quantity in commerce
- 2960 units
- Distribution
- Nationwide.
- Product codes and lots
- Lot number 00VL987139 exp 12/31/2012