Z-2432-2008
Xpect Giardia/Cryptosporidium Kit, Ref 24050520, 20 tests per kit, For In Vitro Diagnostic Use. A rapid test...
Remel Inc / initiated 2008-08-11 / Terminated / root cause: manufacturing process
Remel, Inc began a recall of Xpect Giardia/Cryptosporidium Kit, Ref 24050520, 20 tests per kit, For In Vitro Diagnostic Use. A rapid test for the direct qualitative detection of Giardia... on . The reason FDA records is "Faint grey test lines for Cryptosporidium may be visible which may be interpreted as false positive". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2432-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Remel Inc
- Product
- Xpect Giardia/Cryptosporidium Kit, Ref 24050520, 20 tests per kit, For In Vitro Diagnostic Use. A rapid test for the direct qualitative detection of Giardia and Cryptosporidium antigens in fecal specimens.
- Product code
- MHI Giardia Spp.
- Reason for recall
- Faint grey test lines for Cryptosporidium may be visible which may be interpreted as false positive
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- The recalling firm issued letters dated 6/19/08 flagged "IMPORTANT PRODUCT INFORMATION" which advised their customers of the issue with the faint grey test lines but did not request any action be taken, such as return or destruction of the affected lot numbers. The recalling firm issued a second letter flagged "IMPORTANT MEDICAL DEVICE CUSTOMER ADVISORY NOTICE" dated 8/11/08 and was issued via regular mail on 8/14/08. This letter asked users to review their inventory for the product lots indicated in the letter, discard all remaining units of this product lot, complete and return the Product Inventory Checklist regardless if they have affected product. A "PRODUCT INVENTORY CHECKLIST" was enclosed to report whether the firm had product to discard and the amount to be replaced.
- Quantity in commerce
- 3,974 kits
- Distribution
- Worldwide Distribution --- including USA and countries of Canada and U.K.
- Product codes and lots
- Lot numbers 597374, Exp. 2008.10.08; 607636, Exp. 2008.11.06; 615970, Exp. 2008.12.02; 615971, Exp. 2008.12.02; 621187, Exp. 2008.12.02; 622617, Exp. 2009.04.02; 624618, Exp. 2009.01.02; 635553, Exp. 2009.04.17; 635554, Exp. 2009.04.24; 640609, Exp. 2009.05.01; 654829, Exp. 2009.05.30; and 658878, Exp. 2009.05.30