Z-0209-2009

Iris Sample Processing SafeCrit 40mm Tubes. Catalog number 59-000869-002 H. Iris Sample Processing, 60...

Drummond Scientific Co / initiated 2008-08-01 / Terminated / root cause: manufacturing process

Drummond Scientific Co began a recall of Iris Sample Processing SafeCrit 40mm Tubes. Catalog number 59-000869-002 H. Iris Sample Processing, 60 Glacier Drive, Westwood, MA 02090. Heparin-treated... on . The reason FDA records is "presence of OSCS contaminant". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0209-2009
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Drummond Scientific Co
Product
Iris Sample Processing SafeCrit 40mm Tubes. Catalog number 59-000869-002 H. Iris Sample Processing, 60 Glacier Drive, Westwood, MA 02090. Heparin-treated capillary tube used for spun hematocrit packed cell volume (PCV) determination.
Product code
GIO Tube, Collection, Capillary Blood
Reason for recall
presence of OSCS contaminant
Root cause tag
manufacturing process
FDA root cause
Material/Component Contamination
Action
The recalling firm issued recall letters dated 8/1/08 via certified mail to the label owners informing them of the problem and the need to notified their customers.
Quantity in commerce
4000 vials
Distribution
Nationwide Distribution --- including states of FL and MA.
Product codes and lots
Lot numbers 8121 exp 6/10, 8142 exp 7/10, & 8165 exp 8/10 and Catalog number 59-000857-002 G - Lot number 8200 exp 9/10