Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2418-2008

Zimmer Durom cup, Metasul Durom acetabular component, uncemented, 50/44, code J, Zimmer, Winterthur...

Zimmer GmbH / initiated 2008-07-22 / Terminated / root cause: labeling and UDI

Zimmer Inc. began a recall of Zimmer Durom cup, Metasul Durom acetabular component, uncemented, 50/44, code J, Zimmer, Winterthur, Switzerland; REF 01.00214.150. The product is component of... on . The reason FDA records is "Instructions for use/surgical technique instructions are inadequate". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-2418-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Zimmer GmbH
Product
Zimmer Durom cup, Metasul Durom acetabular component, uncemented, 50/44, code J, Zimmer, Winterthur, Switzerland; REF 01.00214.150. The product is component of the Metasul LDH Head System intended for Noninflammatory degenerative joint disease. (Hip replacement)
Product code
KWA Prosthesis, Hip, Semi-Constrained (Metal Uncemented Acetabular Component)
Reason for recall
Instructions for use/surgical technique instructions are inadequate.
Root cause tag
labeling and UDI
FDA root cause
Labeling False and Misleading
Action
A press release issued on 7/23/08. A letter was sent to implanting surgeons on 7/22/08 advising them of the problem and instructing them to stop implanting these devices until they receive training. Zimmer suspended all marketing and distribution in the U.S. pending completion of user training and revision of the instructions for surgical technique. Zimmer will conduct a removal of all copies of the current surgical technique document and replace it with revised surgical technique document(s). Contact Zimmer, Inc. at 1-866-946-5633 for assistance. Update: A dear surgeon letter was mailed 8/4/08 and provided patient management guidelines. A correction update was sent 8/16/08 with updated surgical technique and instructions for use documents. A dear surgeon letter was mailed 9/25/08 providing additional information regarding the training program.
Quantity in commerce
19,014 of all products.
Distribution
Nationwide.
Product codes and lots
All units.