Z-2424-2008
Zimmer Durom cup, Metasul Durom acetabular component, uncemented, 62/56, code V, Zimmer, Winterthur...
Zimmer GmbH / initiated 2008-07-22 / Terminated / root cause: labeling and UDI
Zimmer Inc. began a recall of Zimmer Durom cup, Metasul Durom acetabular component, uncemented, 62/56, code V, Zimmer, Winterthur, Switzerland; REF 01.00214.162. The product is component of... on . The reason FDA records is "Instructions for use/surgical technique instructions are inadequate". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-2424-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Zimmer GmbH
- Product
- Zimmer Durom cup, Metasul Durom acetabular component, uncemented, 62/56, code V, Zimmer, Winterthur, Switzerland; REF 01.00214.162. The product is component of the Metasul LDH Head System intended for Noninflammatory degenerative joint disease. (Hip replacement)
- Product code
- KWA Prosthesis, Hip, Semi-Constrained (Metal Uncemented Acetabular Component)
- Reason for recall
- Instructions for use/surgical technique instructions are inadequate.
- Root cause tag
- labeling and UDI
- FDA root cause
- Labeling False and Misleading
- Action
- A press release issued on 7/23/08. A letter was sent to implanting surgeons on 7/22/08 advising them of the problem and instructing them to stop implanting these devices until they receive training. Zimmer suspended all marketing and distribution in the U.S. pending completion of user training and revision of the instructions for surgical technique. Zimmer will conduct a removal of all copies of the current surgical technique document and replace it with revised surgical technique document(s). Contact Zimmer, Inc. at 1-866-946-5633 for assistance. Update: A dear surgeon letter was mailed 8/4/08 and provided patient management guidelines. A correction update was sent 8/16/08 with updated surgical technique and instructions for use documents. A dear surgeon letter was mailed 9/25/08 providing additional information regarding the training program.
- Quantity in commerce
- 19,014 of all products.
- Distribution
- Nationwide.
- Product codes and lots
- All units.