Z-2439-2008

500 mL InfoV.A.C. Canisters (with Gel); Part #M8275063/5 (5 canister pack) and Part #M8275063/10 (10 canister...

KCI USA, Inc. / initiated 2008-07-29 / Terminated / root cause: manufacturing process

KCI USA, Inc. began a recall of 500 mL InfoV.A.C. Canisters (with Gel); Part #M8275063/5 (5 canister pack) and Part #M8275063/10 (10 canister pack); products are single use and labeled as... on . The reason FDA records is "The port on the 500mL InfoV.A.C. Canister that connects to the suction pump tubing was found to be partially or fully occluded with plastic". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2439-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
KCI USA, Inc.
Product
500 mL InfoV.A.C. Canisters (with Gel); Part #M8275063/5 (5 canister pack) and Part #M8275063/10 (10 canister pack); products are single use and labeled as STERILE; distributed by Kinetic Concepts, Inc., San Antonio, TX.
Product code
JCX Apparatus, Suction, Ward Use, Portable, Ac-Powered
Reason for recall
The port on the 500mL InfoV.A.C. Canister that connects to the suction pump tubing was found to be partially or fully occluded with plastic.
Root cause tag
manufacturing process
FDA root cause
Equipment maintenance
Action
Urgent - Medical Device Recall letters were sent within the US on 07/29/08 via UPS next day delivery service and a return receipt was maintained to demonstrate delivery of the recall to all consignees. Additionally, firm is sending representatives into accounts which received more than three sales units of canisters and to all foreign consignee locations to facilitate quarantine and replacement of affected stock. Consignees asked to confirm that adequate flow is being delivered and that the dressing is being drawn down for those patients currently using the InfoV.A.C. Canisters. Also, consignees are to locate and quarantine all packaged 500 ml canisters without the green inspection stickers from within the affected lot numbers in their inventory. A KCI Rep will contact them to help facilitate this process. Consignees are to complete and return the Recall Acknowledgement and Inventory Accounting Form. KCI will replace any affected product at their facility and arrange for the return of any affected product.
Quantity in commerce
150,125 canisters.
Distribution
Worldwide Distribution --- including USA and countries of Austria, Belgium, Denmark, France, Germany, Netherlands, Sweden, and the UK.
Product codes and lots
For M8275063/5: 20064392, 20069119, 20069125, 20069126, 20069279, 20069279.18308F, 20075082, and 20079173  For M8275063/10: 20046268, 20046268.18908A, 20046270, 20063933, 20064393, 20064394, 20064395, 20064753, 20064753.14408B, 20067300, 20067300.14408B, 20067301, 20067302, 20067303, 20067304, 20067305, 20069120, 20069121, 20069273, 20069273.14408B; 20069274, 20069275, 20069275.18908A, 20069277, 20069277.18908A, 20069278, 20069278.18908A, 20069280, 20075084, 20075085, 20075086, 20075262, 20075263, 20075264, 20075265, 20075266, 20075268, 20075269, 20079174, 20079175, 20079176, 20079177, 20079178, 20079179, 20079180, 20079181, 20079182, 20079183, 20079184, 20079185, 20079186, 20079187, 20079188, 20089680, 20089681, 20089682, 20089712, 20089713, 20089749, 20089750, 20093372, 20093374, and 20094200