Z-0533-2009

Depuy DELTA Xtend, Standard Humerus PE Cup, Dia 38/+6 STD, sterile, Depuy France; REF 130738206. The device...

Depuy Orthopaedics Inc. / initiated 2008-07-10 / Terminated / root cause: design

Depuy Orthopaedics, Inc. began a recall of Depuy DELTA Xtend, Standard Humerus PE Cup, Dia 38/+6 STD, sterile, Depuy France; REF 130738206. The device is used as a shoulder implant on . The reason FDA records is "Compressed air may become trapped between the cup and the stem, preventing the tapered walls from engaging and causing the cup to feel unstable". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0533-2009
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Depuy Orthopaedics Inc.
Product
Depuy DELTA Xtend, Standard Humerus PE Cup, Dia 38/+6 STD, sterile, Depuy France; REF 130738206. The device is used as a shoulder implant.
Product code
KWS Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer Cemented
Reason for recall
Compressed air may become trapped between the cup and the stem, preventing the tapered walls from engaging and causing the cup to feel unstable.
Root cause tag
design
FDA root cause
Device Design
Action
A letter, "Urgent Device Correction" dated July 8, 2008 was sent via certified mail to implanting surgeons advising them of the issues and guidance on how to mitigate the problem. A communication was also sent to all sales representatives and territory offices on July 15, 2008. For questions, please contact Depuy Orthopaedics, Inc. Customer Service at 1-800-473-3789.
Quantity in commerce
1,521 of all devices.
Distribution
Nationwide Distribution.
Product codes and lots
All lots.