Z-0537-2009
Depuy DELTA Xtend, Standard Humeral PE Cup, Dia 42/+6 STD, sterile, Depuy France; REF 130742206. The device...
Depuy Orthopaedics Inc. / initiated 2008-07-10 / Terminated / root cause: design
Depuy Orthopaedics, Inc. began a recall of Depuy DELTA Xtend, Standard Humeral PE Cup, Dia 42/+6 STD, sterile, Depuy France; REF 130742206. The device is used as a shoulder implant on . The reason FDA records is "Compressed air may become trapped between the cup and the stem, preventing the tapered walls from engaging and causing the cup to feel unstable". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0537-2009
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Depuy Orthopaedics Inc.
- Product
- Depuy DELTA Xtend, Standard Humeral PE Cup, Dia 42/+6 STD, sterile, Depuy France; REF 130742206. The device is used as a shoulder implant.
- Product code
- KWS Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer Cemented
- Reason for recall
- Compressed air may become trapped between the cup and the stem, preventing the tapered walls from engaging and causing the cup to feel unstable.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- A letter, "Urgent Device Correction" dated July 8, 2008 was sent via certified mail to implanting surgeons advising them of the issues and guidance on how to mitigate the problem. A communication was also sent to all sales representatives and territory offices on July 15, 2008. For questions, please contact Depuy Orthopaedics, Inc. Customer Service at 1-800-473-3789.
- Quantity in commerce
- 1, 521 of all devices.
- Distribution
- Nationwide Distribution.
- Product codes and lots
- All lots.