Z-0541-2009

MicroScan Synergies Plus Negative Panels, 20 panels per box Medical device designed for use in determining...

Siemens Healthcare Diagnostics Inc. / initiated 2008-07-31 / Terminated

Siemens Microscan began a recall of MicroScan Synergies Plus Negative Panels, 20 panels per box Medical device designed for use in determining quantitative and/or qualitative antimicrobial agent... on . The reason FDA records is "Falsely lowered imipenem MICs may result, leading to major errors in results". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0541-2009
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
MicroScan Synergies Plus Negative Panels, 20 panels per box Medical device designed for use in determining quantitative and/or qualitative antimicrobial agent susceptibility and/or identification to the species level of colonies, grown on solid media, of rapidly growing aerobic and facultative anaerobic gram-negative bacteria. Catalog Nos. B1016-200, B1016-201, B1016-202, B1016-203, B1025-103, B1025-104, B1025-105, B1025-106, B1025-107, B1025-108, B1025-109, B1025-112, and B1027-101
Product code
JWY Manual Antimicrobial Susceptibility Test Systems
Reason for recall
Falsely lowered imipenem MICs may result, leading to major errors in results.
FDA root cause
Other
Action
Consignees were notified on July 31, 2008 by letter (Urgent Device Correction) sent via FedEx overnight. International consignees were notified by Siemens Healthcare Diagnostics associates in affected countries. The letter described the problem and provided users with safety instructions to follow until an update is provided. Contact Siemens Healthcare Diagnostics Inc. at 916-374-3075 for additional assistance.
Quantity in commerce
Total of both Negative and Synergy: 569,375 boxes
Distribution
Worldwide
Product codes and lots
All lots