Z-2466-2008

FOUNDATION KNEE SYSTEM Tibial Fixed Impactor, Revision A and B; Catalog #801-01-449; distributed by Encore...

Encore Medical, LP / initiated 2008-08-04 / Terminated / root cause: design

Encore Medical, Lp began a recall of FOUNDATION KNEE SYSTEM Tibial Fixed Impactor, Revision A and B; Catalog #801-01-449; distributed by Encore Medical, L.P., Austin, TX 78758. The Tibial Fixed... on . The reason FDA records is "Complaints received regarding the fracture of the Nitronic 60 screw on the Tibial Fixed Impactor during surgery". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-2466-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Encore Medical, LP
Product
FOUNDATION KNEE SYSTEM Tibial Fixed Impactor, Revision A and B; Catalog #801-01-449; distributed by Encore Medical, L.P., Austin, TX 78758. The Tibial Fixed Impactor, a part of the FOUNDATION KNEE SYSTEM, is a manual orthopedic surgical instrument intended to lock onto the baseplate to impact the implant.
Product code
JWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer
Reason for recall
Complaints received regarding the fracture of the Nitronic 60 screw on the Tibial Fixed Impactor during surgery.
Root cause tag
design
FDA root cause
Device Design
Action
Recall notification sent via letter starting on 08/04/07. Consignees asked to quarantine and return affected devices. Contact Encore Medical, Lp at 1-512-834-6330 for assistance.
Quantity in commerce
68 units.
Distribution
FL and TX and OUS: Japan.
Product codes and lots
All Revision A and Revision B Impactors.