Z-2466-2008
FOUNDATION KNEE SYSTEM Tibial Fixed Impactor, Revision A and B; Catalog #801-01-449; distributed by Encore...
Encore Medical, LP / initiated 2008-08-04 / Terminated / root cause: design
Encore Medical, Lp began a recall of FOUNDATION KNEE SYSTEM Tibial Fixed Impactor, Revision A and B; Catalog #801-01-449; distributed by Encore Medical, L.P., Austin, TX 78758. The Tibial Fixed... on . The reason FDA records is "Complaints received regarding the fracture of the Nitronic 60 screw on the Tibial Fixed Impactor during surgery". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-2466-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Encore Medical, LP
- Product
- FOUNDATION KNEE SYSTEM Tibial Fixed Impactor, Revision A and B; Catalog #801-01-449; distributed by Encore Medical, L.P., Austin, TX 78758. The Tibial Fixed Impactor, a part of the FOUNDATION KNEE SYSTEM, is a manual orthopedic surgical instrument intended to lock onto the baseplate to impact the implant.
- Product code
- JWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer
- Reason for recall
- Complaints received regarding the fracture of the Nitronic 60 screw on the Tibial Fixed Impactor during surgery.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Recall notification sent via letter starting on 08/04/07. Consignees asked to quarantine and return affected devices. Contact Encore Medical, Lp at 1-512-834-6330 for assistance.
- Quantity in commerce
- 68 units.
- Distribution
- FL and TX and OUS: Japan.
- Product codes and lots
- All Revision A and Revision B Impactors.