Z-2445-2008

Universal Optical Tracker Fixation, not sterile, Manufactured by Orthosoft, Montreal, Canada, REF 111.031...

Zimmer GmbH / initiated 2007-08-06 / Terminated / root cause: design

Zimmer Inc. began a recall of Universal Optical Tracker Fixation, not sterile, Manufactured by Orthosoft, Montreal, Canada, REF 111.031. Surgical instrument on . The reason FDA records is "Inadequate weld on tracker, knob oversized and poor thread design on attachment screw results in seizing. The instrument may break during use, resulting in surgical delay and an increased risk of infection". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-2445-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Zimmer GmbH
Product
Universal Optical Tracker Fixation, not sterile, Manufactured by Orthosoft, Montreal, Canada, REF 111.031. Surgical instrument.
Product code
HAW Neurological Stereotaxic Instrument
Reason for recall
Inadequate weld on tracker, knob oversized and poor thread design on attachment screw results in seizing. The instrument may break during use, resulting in surgical delay and an increased risk of infection.
Root cause tag
design
FDA root cause
Device Design
Action
Firm's sales representatives were notified via e-mail on 8/6/07 to review their inventory and that of their accounts and to notify the firm of their location so that they could be replaced. Contact the firm by telephone at 574-267-6131 for additional information concerning this recall.
Quantity in commerce
39
Distribution
Nationwide Distribution including states of California, Florida and Washington.
Product codes and lots
Lots 050433, BP050199, BP050258 and BP050350.