Z-1328-2009
Stat Profile Critical Care Xpress (CCX and CCX+), CCX Creatinine Membrane Kit for CCX analyzer Catalog #...
Nova Biomedical Corporation / initiated 2008-08-11 / Terminated / root cause: manufacturing process
Nova Biomedical Corporation began a recall of Stat Profile Critical Care Xpress (CCX and CCX+), CCX Creatinine Membrane Kit for CCX analyzer Catalog # 35238 Intended for in vitro diagnostic use by health... on . The reason FDA records is "Erroneous, low creatinine results on CCX analyzer". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1328-2009
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Nova Biomedical Corporation
- Product
- Stat Profile Critical Care Xpress (CCX and CCX+), CCX Creatinine Membrane Kit for CCX analyzer Catalog # 35238 Intended for in vitro diagnostic use by health care professionals and for point-of-care usage in the quantitative determination of pH, PCO2, PO2, SO2%, Hematocrit (Hct), Ca++, total hemoglobin (tHb), Oxyhemoglobin (O2Hb), Carboxyhemoglobin (COHb), Methemoglobin (MetHb), reduced hemoglobin (HHb), Oxygen content (O2Ct) and Oxygen capacity (O2Cap) in heparinized whole blood; Na+, K+, CL-; Ca++, Mg++, Glucose heparinized whole blood, serum, or plasma.
- Product code
- CGL Electrode, Ion Based, Enzymatic, Creatinine
- Reason for recall
- Erroneous, low creatinine results on CCX analyzer
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Initial: USA and Canada - Customer Advisory Notice notification by phone contact with key customer contact facilitated using a Field Correction Script. This contact will be documented within the Nova Biomedical Technical Support database (Pivotal) and on a Field Correction Checklist. All affected inventory will be replaced at customer sites. International - Customer Advisory Notice and Faxback Form being communicated and sent to Nova Biomedical Subsidiaries and Distributors by three methods of delivery (Email, Fax and Phone Call). Follow-up: USA and Canada - Sending the Customer Advisory Notice by email and/or fax to the key customer contact from the initial phone contact. International - Sending the Customer Advisory Notice by email and/or fax to the key contact.
- Quantity in commerce
- 1230
- Distribution
- Worldwide Distribution -- USA, Canada, Germany, France, Czech Republic, Taiwan, Japan, Thailand, Italy, Hungary, India, UK, Ireland, Venezuela, Brazil, and Hong Kong.
- Product codes and lots
- Lots 806050, 806182, 806242, 806567, 806721, 806752, and 807246.