Z-1328-2009

Stat Profile Critical Care Xpress (CCX and CCX+), CCX Creatinine Membrane Kit for CCX analyzer Catalog #...

Nova Biomedical Corporation / initiated 2008-08-11 / Terminated / root cause: manufacturing process

Nova Biomedical Corporation began a recall of Stat Profile Critical Care Xpress (CCX and CCX+), CCX Creatinine Membrane Kit for CCX analyzer Catalog # 35238 Intended for in vitro diagnostic use by health... on . The reason FDA records is "Erroneous, low creatinine results on CCX analyzer". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1328-2009
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Nova Biomedical Corporation
Product
Stat Profile Critical Care Xpress (CCX and CCX+), CCX Creatinine Membrane Kit for CCX analyzer Catalog # 35238 Intended for in vitro diagnostic use by health care professionals and for point-of-care usage in the quantitative determination of pH, PCO2, PO2, SO2%, Hematocrit (Hct), Ca++, total hemoglobin (tHb), Oxyhemoglobin (O2Hb), Carboxyhemoglobin (COHb), Methemoglobin (MetHb), reduced hemoglobin (HHb), Oxygen content (O2Ct) and Oxygen capacity (O2Cap) in heparinized whole blood; Na+, K+, CL-; Ca++, Mg++, Glucose heparinized whole blood, serum, or plasma.
Product code
CGL Electrode, Ion Based, Enzymatic, Creatinine
Reason for recall
Erroneous, low creatinine results on CCX analyzer
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Initial: USA and Canada - Customer Advisory Notice notification by phone contact with key customer contact facilitated using a Field Correction Script. This contact will be documented within the Nova Biomedical Technical Support database (Pivotal) and on a Field Correction Checklist. All affected inventory will be replaced at customer sites. International - Customer Advisory Notice and Faxback Form being communicated and sent to Nova Biomedical Subsidiaries and Distributors by three methods of delivery (Email, Fax and Phone Call). Follow-up: USA and Canada - Sending the Customer Advisory Notice by email and/or fax to the key customer contact from the initial phone contact. International - Sending the Customer Advisory Notice by email and/or fax to the key contact.
Quantity in commerce
1230
Distribution
Worldwide Distribution -- USA, Canada, Germany, France, Czech Republic, Taiwan, Japan, Thailand, Italy, Hungary, India, UK, Ireland, Venezuela, Brazil, and Hong Kong.
Product codes and lots
Lots 806050, 806182, 806242, 806567, 806721, 806752, and 807246.