Z-2446-2008

Caption Disposable Platelet Concentrator Kit, Ref 71178200, 1 each, Sterile, Smith & Nephew, Inc. Memphis, TN...

Smith & Nephew, Inc. / initiated 2008-08-05 / Terminated / root cause: packaging

Smith & Nephew Inc began a recall of Caption Disposable Platelet Concentrator Kit, Ref 71178200, 1 each, Sterile, Smith & Nephew, Inc. Memphis, TN 38116 on . The reason FDA records is "Product contained a syringe whose package seal integrity can be adversely affected when the product is exposed to low atmospheric pressure. This may occur during shipping". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-2446-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Smith & Nephew, Inc.
Product
Caption Disposable Platelet Concentrator Kit, Ref 71178200, 1 each, Sterile, Smith & Nephew, Inc. Memphis, TN 38116.
Product code
FMF Syringe, Piston
Reason for recall
Product contained a syringe whose package seal integrity can be adversely affected when the product is exposed to low atmospheric pressure. This may occur during shipping.
Root cause tag
packaging
FDA root cause
Other
Action
Prior to the dissemination of the official notices, the firm notified personnel at our field offices of the BD syringe recall and advised that a subsequent recall of our Caption devices would follow. High volume users were contacted via phone on July 30th and the 31st and personnel at all affected field offices were notified via email on July 30, 2008 with an Urgent Medical Device Recall Letter. The firm requedsted that usere immediately cease distribution, examin the stock on hand and the stock of the accounts that have been suppplied and remove any affected ecices from all inventory. A verification notification was included with letter
Quantity in commerce
364 units
Distribution
Nationwide Distribution
Product codes and lots
Lot Nos.: 07EQT0003, 07EQT0004, 07EQT0005, 07JQT0003, 07KQT0002, 63207107, 07HM12429, and 07JM15582.