Z-2447-2008
Caption Disposable Platelet Concentrator Kit, Ref 71178202, 1 each, Sterile, Smith & Nephew, Inc. Memphis, TN...
Smith & Nephew, Inc. / initiated 2008-08-05 / Terminated / root cause: packaging
Smith & Nephew Inc began a recall of Caption Disposable Platelet Concentrator Kit, Ref 71178202, 1 each, Sterile, Smith & Nephew, Inc. Memphis, TN 38116 on . The reason FDA records is "Product contained a syringe whose package seal integrity can be adversely affected when the product is exposed to low atmospheric pressure. This may occur during shipping". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-2447-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Smith & Nephew, Inc.
- Product
- Caption Disposable Platelet Concentrator Kit, Ref 71178202, 1 each, Sterile, Smith & Nephew, Inc. Memphis, TN 38116
- Product code
- FMF Syringe, Piston
- Reason for recall
- Product contained a syringe whose package seal integrity can be adversely affected when the product is exposed to low atmospheric pressure. This may occur during shipping.
- Root cause tag
- packaging
- FDA root cause
- Other
- Action
- Prior to the dissemination of the official notices, the firm notified personnel at our field offices of the BD syringe recall and advised that a subsequent recall of our Caption devices would follow. High volume users were contacted via phone on July 30th and the 31st and personnel at all affected field offices were notified via email on July 30, 2008 with an Urgent Medical Device Recall Letter. The firm requedsted that usere immediately cease distribution, examin the stock on hand and the stock of the accounts that have been suppplied and remove any affected ecices from all inventory. A verification notification was included with letter
- Quantity in commerce
- 568 units
- Distribution
- Nationwide Distribution
- Product codes and lots
- Lot Nos.: 08EQT0001, 08DQT0001, 08CQT0003, 08BQT0002, and 08AQT0003