Z-2457-2008
Cholestech LDX High Sensitivity C-Reactive Protein (hs-CRP) Test Cassette. The Cholestech LDX high...
Cholestech Corp. / initiated 2006-08-09 / Terminated / root cause: design
Cholestech Corp began a recall of Cholestech LDX High Sensitivity C-Reactive Protein (hs-CRP) Test Cassette. The Cholestech LDX high sensitivity C-Reactive Protein (hs-CRP) is an in vitro... on . The reason FDA records is "Test results are high, outside of control material upper limit specification. Results could be as high as 15%". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-2457-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Cholestech Corp.
- Product
- Cholestech LDX High Sensitivity C-Reactive Protein (hs-CRP) Test Cassette. The Cholestech LDX high sensitivity C-Reactive Protein (hs-CRP) is an in vitro diagnostic test for the quantitative determination of C-reactive protein in whole blood or serum. Measurement of CRP is useful as an aid in the detection and evaluation of infection, tissue injury, inflammatory disorders and associated diseases. The test cassettes are for use with the LDX analyzer.
- Product code
- DCK C-Reactive Protein, Antigen, Antiserum, And Control
- Reason for recall
- Test results are high, outside of control material upper limit specification. Results could be as high as 15%.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Cholestech provided its three distributors a notification letter/response form with instructions to distribute the letter to their customers/end users dated August 9, 2006. The letter instructed customers to discontinue use of the lot of cassettes, return remaining cassettes to Cholestech through the Technical Service Representative and to complete and return the response form. The end users returned the signed letters directly to Cholestech so that effectiveness checks could be completed.
- Quantity in commerce
- 2, 910 cassettes
- Distribution
- Nationwide Distribution --- including states of Virginia, Texas and North Carolina, Ohio, Arkansas, Washington, Florida , California and Michigan.
- Product codes and lots
- Catalog #12-807, Lot #C17-6434