Z-2460-2008

Cholestech GDX A1C Test Cartridge. The Cholestech GDX A1C Test is an in-vitro diagnostic test that measures...

Cholestech Corp. / initiated 2006-08-03 / Terminated / root cause: manufacturing process

Cholestech Corp began a recall of Cholestech GDX A1C Test Cartridge. The Cholestech GDX A1C Test is an in-vitro diagnostic test that measures hemoglobin A1c (HbA1c) on . The reason FDA records is "Cholestech Corp. was notified by the foreign manufacturer of the A1C test cartridge that stability data indicated that the product would not meet performance claims through the end of its shelf life". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2460-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Cholestech Corp.
Product
Cholestech GDX A1C Test Cartridge. The Cholestech GDX A1C Test is an in-vitro diagnostic test that measures hemoglobin A1c (HbA1c).
Product code
LCP Assay, Glycosylated Hemoglobin
Reason for recall
Cholestech Corp. was notified by the foreign manufacturer of the A1C test cartridge that stability data indicated that the product would not meet performance claims through the end of its shelf life.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
A customer notification letter was sent to Cholestech consignees who had received product from this lot. The letter dated 8/3/2006 instructed customers to discontinue use of the lot of product, destroy product according to in-house standard procedures and contact Cholestech Technical Services for replacement. It also instructed to provide a copy of the letter to other users if the lot of product was further distributed. A response form was included with instructions to complete and return. Contact Cholestech Corp at 1-510-781-3505 for assistance.
Quantity in commerce
9,324 Cartridges
Distribution
U.S. state of Colorado. OUS: India, Slovenia, Germany, Tunisia, Taiwan, Canada, Philippines, South Africa and Mexico
Product codes and lots
Catalog #12-658, Lot #066T28