Ev3, Inc began a recall of ev3" 55442-2920, sterile EO, single Use, Visi-Pro Balloon-Expandable Peripheral Stent System. Model Numbers: PXB35-05-12-080, PXB35-06-17-135, PXB35-06-27-080... on . The reason FDA records is "ev3, Inc. has determined that specific lots of the Visi-Pro" BE Stent system may not have adequate securement of the stent to the delivery system and may have a larger diameter/profile than intended". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1547-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Medtronic, Inc.
- Product
- ev3" 55442-2920, sterile EO, single Use, Visi-Pro Balloon-Expandable Peripheral Stent System. Model Numbers: PXB35-05-12-080, PXB35-06-17-135, PXB35-06-27-080, PXB35-06-57-135, PXB35-07-37-080 ,PXB35-07-57-080, PXB35-08-17-080, PXB35-08-17-135, PXB35-08-37-080, PXB35-08-57-080, PXB35-09-17-135 ,PXB35-09-27-080, PXB35-09-27-135, PXB35-09-57-080, PXB35-10-27-135, and PXB35-10-37-080. The Endovascular Company , 4800 Nathan Lane North, Plymouth, MN
- Product code
- FGE Stents, Drains And Dilators For The Biliary Ducts
- Reason for recall
- ev3, Inc. has determined that specific lots of the Visi-Pro" BE Stent system may not have adequate securement of the stent to the delivery system and may have a larger diameter/profile than intended.
- Root cause tag
- manufacturing process
- FDA root cause
- Employee error
- Action
- Consignees were sent a "Medical Device Recall" letter dated 7/1/08. The letter described the product, problem and action to be taken. The letter also requested to complete and return The Field Action Reconciliation Form. For further information, please contact Customer Service at 800-716-6700.
- Quantity in commerce
- 66
- Distribution
- Worldwide Distribution including states of AL,AZ, CA, CO, CT, FL, IL, IN, IA, KS, KY, LA,ME, MA, MI, MN, MO, MT, NJ, NY, NC, OH, OR, PA, RI, SC, TN, TX, WV, WI, and VA. Counties of SWEDEN, TUNISIA, SWITZERLAND, GERMANY, ITALY, SLOVENIA, FRANCE, BELGIUM, UNITED KINGDOM, TURKEY, FINLAND, LITHUANIA, and NORWAY.
- Product codes and lots
- Lot Numbers: 3273766, 3273767, 3509137, 4218926, 4361370, 4472112, 4941116, 5012047, 5282885, and 5834997.