Sonoline G50/G60 Ultrasound System, Software versions 1.0.0 to 1.5.0b. The product is a general purpose...
Siemens Medical Solutions USA, Inc. began a recall of Sonoline G50/G60 Ultrasound System, Software versions 1.0.0 to 1.5.0b. The product is a general purpose, mobile software controlled diagnostic ultrasound system on . The reason FDA records is "As a result of bugs and calculation errors, the system may display incorrect or fail to display mechanical Index and thermal index values". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-2468-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Product
- Sonoline G50/G60 Ultrasound System, Software versions 1.0.0 to 1.5.0b. The product is a general purpose, mobile software controlled diagnostic ultrasound system.
- Product code
- IYN System, Imaging, Pulsed Doppler, Ultrasonic
- Reason for recall
- As a result of bugs and calculation errors, the system may display incorrect or fail to display mechanical Index and thermal index values.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- The firm initiated its field correction on 08/05/2008. The firm's service personnel will make site visits to correct the problem and install the revised software (version 1.5.0c). Contact Siemens Medical Solutions USA, Inc. at 1-650-694-5993 for assistance.
- Quantity in commerce
- 2541 units
- Distribution
- Worldwide
- Product codes and lots
- Model numbers 07482800, 07482818, 07482826, 07478139, 07478147, and 07482479