Z-2468-2008

Sonoline G50/G60 Ultrasound System, Software versions 1.0.0 to 1.5.0b. The product is a general purpose...

Siemens Medical Solutions USA, Inc. / initiated 2008-08-05 / Terminated / root cause: software

Siemens Medical Solutions USA, Inc. began a recall of Sonoline G50/G60 Ultrasound System, Software versions 1.0.0 to 1.5.0b. The product is a general purpose, mobile software controlled diagnostic ultrasound system on . The reason FDA records is "As a result of bugs and calculation errors, the system may display incorrect or fail to display mechanical Index and thermal index values". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-2468-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
Sonoline G50/G60 Ultrasound System, Software versions 1.0.0 to 1.5.0b. The product is a general purpose, mobile software controlled diagnostic ultrasound system.
Product code
IYN System, Imaging, Pulsed Doppler, Ultrasonic
Reason for recall
As a result of bugs and calculation errors, the system may display incorrect or fail to display mechanical Index and thermal index values.
Root cause tag
software
FDA root cause
Software design
Action
The firm initiated its field correction on 08/05/2008. The firm's service personnel will make site visits to correct the problem and install the revised software (version 1.5.0c). Contact Siemens Medical Solutions USA, Inc. at 1-650-694-5993 for assistance.
Quantity in commerce
2541 units
Distribution
Worldwide
Product codes and lots
Model numbers 07482800, 07482818, 07482826, 07478139, 07478147, and 07482479