Z-2469-2008

Dideco, D901 Lilliput 1, Close Phisio, Sterile EO, REF 05252, Manufacturer: Sorin Group Italia, Via Statale...

LivaNova USA, Inc. / initiated 2008-08-08 / Terminated / root cause: labeling and UDI

Sorin Group USA, Inc. began a recall of Dideco, D901 Lilliput 1, Close Phisio, Sterile EO, REF 05252, Manufacturer: Sorin Group Italia, Via Statale 12 Nord, 86, 41037 Mirandola (MO) (Italy). The... on . The reason FDA records is "Under certain conditions, venous reservoir on oxygenator may allow air into the system". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-2469-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
LivaNova USA, Inc.
Product
Dideco, D901 Lilliput 1, Close Phisio, Sterile EO, REF 05252, Manufacturer: Sorin Group Italia, Via Statale 12 Nord, 86, 41037 Mirandola (MO) (Italy). The product is intended for use in cardiopulmonary bypass circuits as a device to replace the function of the lungs in order to control the arterial/venous temperature and as a venous blood reservoir
Product code
DTZ Oxygenator, Cardiopulmonary Bypass
Reason for recall
Under certain conditions, venous reservoir on oxygenator may allow air into the system.
Root cause tag
labeling and UDI
FDA root cause
Labeling False and Misleading
Action
The firm sent out a notification letter, by certified letter, on Aug 8,2008. Instructions on how to prevent air from entering the system were provided. Sorin Group USA, Inc. can reached at 1-303-467-6306.
Quantity in commerce
35 units
Distribution
NC
Product codes and lots
Lot No: 0501210038, and 0710290112.