Z-2654-2010
Laparoscopic Ultrasound Transducer (LAP Transducer) is used on the SonoSite MicroMaxx ultrasound system. The...
FUJIFILM SonoSite,Inc. / initiated 2007-11-02 / Terminated / root cause: manufacturing process
Sonosite, Inc. began a recall of Laparoscopic Ultrasound Transducer (LAP Transducer) is used on the SonoSite MicroMaxx ultrasound system. The LAP Transducer label is labeled in part... on . The reason FDA records is "Sterilization of the SonoSite LAP surgical Transducer with the Sterrad NX System may result in damage to the transducer". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2654-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- FUJIFILM SonoSite,Inc.
- Product
- Laparoscopic Ultrasound Transducer (LAP Transducer) is used on the SonoSite MicroMaxx ultrasound system. The LAP Transducer label is labeled in part: "MicroMaxx LAP...Manufactured for SonoSite, Inc., Bothell, WA 98021".
- Product code
- IYN System, Imaging, Pulsed Doppler, Ultrasonic
- Reason for recall
- Sterilization of the SonoSite LAP surgical Transducer with the Sterrad NX System may result in damage to the transducer.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- On November 2, 2007, SonoSite sent out the Important Information Regarding Sterrad NX Incompatibility letter and the Technical Bulletin 005/07 Rev A to their consignees. The letter advises the consignees that SonoSite has recently found that sterilization of the SonoSite SLT and LAP surgical ultrasound transducers with the Sterrad NX system may result in damage to the transducer. The Technical Bulletin 005/07 Rev B dated November 2, 2007 informs the consignees that the damage has been observed primarily on the circuit board in the transducer connector, and cable and laparoscope damage may also be occurring. All damages have occurred after processing using the Sterrad NX systems in its "Advanced" cycle and the potential damage from the use of "normal" cycle is being evaluated by the firm. Consignees should call the Technical Support at 1-425-951-1330 or 1-877-657-8118 or e-mail to [email protected] for any questions regarding the recall.
- Quantity in commerce
- 16 units
- Distribution
- Devices were distributed to medical centers and hospitals in AR, CA, IN, KS, MA, MD, MI, NC, NE, NJ, NV, OH, SC, TN, TX, WI, Washington DC and to FOREIGN COUNTRIES Australia, China, France, Germany, Greece, Hong Kong, India, Iraq, Italy, Korea, Republic of, Libyan Arab Jamahiriya, Mexico, Spain, United Arab Emirates, United Kingdom, and United States.
- Product codes and lots
- 16 devices are identified with Serial Numbers: 038VWM, 038FK1, 0379P6, 038G67, 038VWK, 038192, 0379P1, 0379P2, 038FJZ, 038FJW, 0379P0, 0379P5, 037WMJ, 038VWP, 037XT6, and 038153.
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.
