Z-0306-2009
CELL-DYN Sapphire Hematology Analyzer, List numbers 08H00-01, 08H00-03 A multi- parameter, automated...
Abbott Medical / initiated 2008-08-25 / Terminated / root cause: design
Abbott Laboratories began a recall of CELL-DYN Sapphire Hematology Analyzer, List numbers 08H00-01, 08H00-03 A multi- parameter, automated hematology analyzer designed for in vitro diagnostic use... on . The reason FDA records is "Exposed wire resulted in minor electrical shock and burn to the service technician". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0306-2009
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Abbott Medical
- Product
- CELL-DYN Sapphire Hematology Analyzer, List numbers 08H00-01, 08H00-03 A multi- parameter, automated hematology analyzer designed for in vitro diagnostic use in counting and characterizing blood cells.
- Product code
- GKZ Counter, Differential Cell
- Reason for recall
- Exposed wire resulted in minor electrical shock and burn to the service technician
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Recall issued August 22, 2008. Technical service bulletin was issued for all field personnel via internal database. No customer communication is planned as customers are instructed *not* to remove the cover, which is necessary to access the area.
- Quantity in commerce
- 958 units
- Distribution
- Worldwide Distribution --- including USA and countries of Argentina, Australia, Brazil, Canada, Chile, China WFOTC, Colombia, Germany, Hong Kong, Japan, Mexico, New Zealand, Panama, Paraguay, Peru, Singapore, South Korea, Taiwan and Venezuela.
- Product codes and lots
- All pre production units with Serial numbers S-001 to S-026 and all production units with serial numbers 42003AZ-42007AZ, 42010AZ-42030AZ, 42032AZ-42033AZ, 42037AZ-42059AZ, 42061AZ-42062AZ, 42064AZ-42067AZ, 42072AZ-42074AZ, 42076AZ-42081AZ, 42083AZ-42123AZ, 42125AZ-42852AZ, 42854AZ-42944AZ, 42946AZ-42949AZ, 42951AZ, 42953AZ, 42955AZ-42961AZ, and 42963AZ-42964AZ