ConMed Corporation began a recall of VCARE (Vaginal-Cervical-Ahluwalia's-Retractor-Elevator) SMALL, Uterine Manipulator, REF 60-6085-101. Made in USA, CONMED CORPORATION, UTICA, NEW YORK USA 13502 on . The reason FDA records is "The firm received several complaints sighting detachment of the balloon at the distal end of the shaft. It was also noted that the incidence of the forward cup slipping off the shaft and being retained in the vaginal...". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0179-2009
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- ConMed Corporation
- Product
- VCARE (Vaginal-Cervical-Ahluwalia's-Retractor-Elevator) SMALL, Uterine Manipulator, REF 60-6085-101. Made in USA, CONMED CORPORATION, UTICA, NEW YORK USA 13502.
- Product code
- LFK Catheter, Femoral
- Reason for recall
- The firm received several complaints sighting detachment of the balloon at the distal end of the shaft. It was also noted that the incidence of the forward cup slipping off the shaft and being retained in the vaginal canal had increased.
- FDA root cause
- Process change control
- Action
- The domestic recall notices were mailed on 9/08/08 by USP Priority Mail pouches. The international notices were sent via FedEx International Priority on 9/08/08. The letter described the problem, the health risk and requested that customers inventory the product and return them back to the firm. Customers are requested to fax the reply form to the firm. Contact ConMed Corporation at 1-800-765-8375, ext 3237 for assistance.
- Quantity in commerce
- 32,453 (total devices)
- Distribution
- Nationwide, Australia, Belgium, Brazil, Chile, Ecuador, France, and Switzerland
- Product codes and lots
- Lots 0710121x through and including 0804111x [yymmddx where x = shift code 1,2, or 3] . All lots of the recalled devices were manufactured between and including October 12, 2007 and April 11, 2008.