Z-0111-2009

Siemens Ultrasound, Acuson/Sonovista X300 ultrasound systems, software revisions 2.0.1 to 2.0.05, 3.0.01 and...

Siemens Medical Solutions USA, Inc. / initiated 2008-08-19 / Terminated / root cause: manufacturing process

Siemens Medical Solutions USA, Inc. began a recall of Siemens Ultrasound, Acuson/Sonovista X300 ultrasound systems, software revisions 2.0.1 to 2.0.05, 3.0.01 and 3.0.02, manufactured by Siemens Ultrasound on . The reason FDA records is "Thermal Index cranial (TIC) is not displayed for the Neo-Head exam type with the C8-5 transducer. This is a required display to support the alara principle when imaging cranial structure". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0111-2009
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
Siemens Ultrasound, Acuson/Sonovista X300 ultrasound systems, software revisions 2.0.1 to 2.0.05, 3.0.01 and 3.0.02, manufactured by Siemens Ultrasound
Product code
IYN System, Imaging, Pulsed Doppler, Ultrasonic
Reason for recall
Thermal Index cranial (TIC) is not displayed for the Neo-Head exam type with the C8-5 transducer. This is a required display to support the alara principle when imaging cranial structure.
Root cause tag
manufacturing process
FDA root cause
Employee error
Action
Field correction was initiated on 8/19/2008. Under Siemens update program US020/08/S & US021/08/S Siemens service personnel make a site visit, deliver a customer safety advisory notice to the customer and correct the problem by installing the revised Software versions. After the planned field correction is performed at the customer site the Service personnel log that information through Siemens internal web or reporting tools into the Siemens field update database.
Quantity in commerce
72 units
Distribution
Worldwide Distribution --- including USA and countries of France, Colombia, Poland, Taiwan, Saudi Arabia, Austria, Malaysia, India, Czech Republic, Croatia, Argentina, Venezuela, Italy, Philippines, Mauritius, Spain, Bolivia, Germany, Ukrain, Russia, United Kingdom, Brazil, South Africa and Canada.
Product codes and lots
Sonovista: 10133170, Acuson: 10037409 (potentially affected but no volume: 10132987, 10038837). Software revisions 2.0.1 to 2.0.05, 3.0.01 and 3.0.02