Z-0086-2009
Siemens Ultrasound, Acuson/Sonovista X300 ultrasound systems, software revisions Sonovista X300 software...
Siemens Medical Solutions USA, Inc. / initiated 2008-08-19 / Terminated / root cause: software
Siemens Medical Solutions USA, Inc. began a recall of Siemens Ultrasound, Acuson/Sonovista X300 ultrasound systems, software revisions Sonovista X300 software revision 1.0.06 and 1.0.07, Acuson X300 sofware... on . The reason FDA records is "Incorrect value calculations by the device may result in inaccurate aortic stenosis estimates". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0086-2009
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Product
- Siemens Ultrasound, Acuson/Sonovista X300 ultrasound systems, software revisions Sonovista X300 software revision 1.0.06 and 1.0.07, Acuson X300 sofware revision 1.0.08 ultrasound system with onscreen display.
- Product code
- IYN System, Imaging, Pulsed Doppler, Ultrasonic
- Reason for recall
- Incorrect value calculations by the device may result in inaccurate aortic stenosis estimates.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Siemens initiated a Field Correction 08/19/2008. Service personnel will conduct a site visit and distribute a Customer Safety Advisory Notice, correct the problem and install a revised software version.
- Quantity in commerce
- 169 units total
- Distribution
- Worldwide Distribution: USA, Singapore, Germany, Denmark, Romania, Sweden, Poland, Portugal, Australia, Chile, India, Spain, Thailand, Malaysia, South Africa, Hungary, Saudi Arabia, Qatar, Peru, Italy, Azerbaijan, Greenland, Austria, Venezuela, Puerto Rico, Brazil and Japan.
- Product codes and lots
- Sonovista: 10133170 (Japan only) Acuson: 10037409 (potentially affected but no volume: 10132987, 10038837). Software revisions 1.0.06 and 1.0.07 for Sonovista X300 and 1.0.08 for Acuson X300