Z-0114-2009
Safety Lancet for capillary blood sampling, 1.6 mm x 28G, Mini, Sarstedt. (Blood lancet)
Sarstedt ApS / initiated 2008-08-27 / Terminated / root cause: manufacturing process
Sarstedt Inc began a recall of Safety Lancet for capillary blood sampling, 1.6 mm x 28G, Mini, Sarstedt. (Blood lancet) on . The reason FDA records is "The product from this lot may not retract the lancet into the safe position inside the lancet body after the trigger is actuated". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0114-2009
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Sarstedt ApS
- Product
- Safety Lancet for capillary blood sampling, 1.6 mm x 28G, Mini, Sarstedt. (Blood lancet)
- Product code
- FMK Single Use Only Blood Lancet With An Integral Sharps Injury Prevention Feature
- Reason for recall
- The product from this lot may not retract the lancet into the safe position inside the lancet body after the trigger is actuated.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- The consignee was notified via letter that was hand delivered on 08/27/2008. All of the distributed product was picked up from the consignee and returned to Sarstedt, Inc. None of the product had been used. All action was completed on 8/27/2008. Contact Sarstedt at 1-828-465-4000 for assistance.
- Quantity in commerce
- 1000 pieces
- Distribution
- NC
- Product codes and lots
- Product #85.1015.050, Lot #M40C219A7