Z-0134-2009

ANGIOSTAR Plus X-Ray System, Model number 6379718

Siemens Medical Solutions USA, Inc. / initiated 2008-08-11 / Terminated / root cause: software

Siemens Medical Solutions USA, Inc began a recall of ANGIOSTAR Plus X-Ray System, Model number 6379718 on . The reason FDA records is "Unintended movement of the system table and/or C arm". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0134-2009
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
ANGIOSTAR Plus X-Ray System, Model number 6379718.
Product code
IZI System, X-Ray, Angiographic
Reason for recall
Unintended movement of the system table and/or C arm.
Root cause tag
software
FDA root cause
Software design
Action
Siemens Medical Solutions issued a letter to consignees on 8/11/2008 to inform them of the problem and a workaround should the issue occur prior to software revision.
Quantity in commerce
114 units
Distribution
Nationwide.
Product codes and lots
Model number 6379718. Serial numbers 1005, 1006, 1011, 1016, 1023, 1033, 1049, 1057, 1059, 1061, 1067, 1069, 1073, 1081, 1082, 1092, 1093, 1095, 1099, 1100, 1101, 1108, 1109, 1113, 1119, 1120, 1121, 1122, 1123, 1127, 1130, 1131, 1134, 1137, 1140, 1144, 1147, 1153, 1155, 1156, 1157, 1159, 1162, 1168, 1169, 1170, 1171, 1172, 1181, 1184, 1186, 1187, 1207, 1210, 1216, 1221, 1223, 1235, 1236, 1241, 1244, 1247, 1248, 1250, 1260, 1261, 1264, 1265, 1267, 1268, 1271, 1272, 1273, 1274, 1275, 1276, 1278, 1284, 1286, 1288, 1290, 1297, 1306, 1308, 1309, 1311, 1313, 1320, 1321, 1335, 1337, 1338, 1345, 1351, 1352, 1354, 1356, 1366, 1367, 1368, 1369, 1371, 1373, 1376, 1378, 1382, 1383, 1385, 1386, 1387, 1388, 1389, 1395, and 9004.