Z-0022-2009
BD Visitec High Viscosity Injector, 4mm Ref: 585173. The product is intended for Vitroretinal Surgery
Becton Dickinson & Co / initiated 2008-09-15 / Terminated / root cause: manufacturing process
Becton Dickinson and Company began a recall of BD Visitec High Viscosity Injector, 4mm Ref: 585173. The product is intended for Vitroretinal Surgery on . The reason FDA records is "Product labeled as 4 mm High Viscosity Injector tip contains a 6 mm tip". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0022-2009
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Becton Dickinson & Co
- Product
- BD Visitec High Viscosity Injector, 4mm Ref: 585173. The product is intended for Vitroretinal Surgery.
- Product code
- HMX Cannula, Ophthalmic
- Reason for recall
- Product labeled as 4 mm High Viscosity Injector tip contains a 6 mm tip.
- Root cause tag
- manufacturing process
- FDA root cause
- Employee error
- Action
- BD contacted accounts by telephone on 9/15/08 requesting users to discard units and complete the response form. Contact Becton Dickinson and Co. at 1-781-906-7950 for assistance.
- Quantity in commerce
- 1485 units
- Distribution
- AZ, CA, CO, IL, OH, FL, PA , TN, TX,and VA Foreign: Australia and Japan.
- Product codes and lots
- Lot Number: 8018591