Z-0143-2009

BICOR HS X-Ray System, Model number 6005776

Siemens Medical Solutions USA, Inc. / initiated 2008-08-11 / Terminated / root cause: software

Siemens Medical Solutions USA, Inc began a recall of BICOR HS X-Ray System, Model number 6005776 on . The reason FDA records is "Unintended movement of the system table and/or C arm". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0143-2009
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
BICOR HS X-Ray System, Model number 6005776
Product code
IZI System, X-Ray, Angiographic
Reason for recall
Unintended movement of the system table and/or C arm.
Root cause tag
software
FDA root cause
Software design
Action
Siemens Medical Solutions issued a letter to consignees on 8/11/2008 to inform them of the problem and a workaround should the issue occur prior to software revision.
Quantity in commerce
2 UNITS
Distribution
Nationwide.
Product codes and lots
Model number 6005776 - Serial numbers: 3016 and 3026