Z-0148-2009
NEUROSTAR TOP X-Ray System, Model numbers 6258425 and 6379668
Siemens Medical Solutions USA, Inc. / initiated 2008-08-11 / Terminated / root cause: software
Siemens Medical Solutions USA, Inc began a recall of NEUROSTAR TOP X-Ray System, Model numbers 6258425 and 6379668 on . The reason FDA records is "Unintended movement of the system table and/or C arm". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0148-2009
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Product
- NEUROSTAR TOP X-Ray System, Model numbers 6258425 and 6379668.
- Product code
- IZI System, X-Ray, Angiographic
- Reason for recall
- Unintended movement of the system table and/or C arm.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Siemens Medical Solutions issued a letter to consignees on 8/11/2008 to inform them of the problem and a workaround should the issue occur prior to software revision.
- Quantity in commerce
- 57 units
- Distribution
- Nationwide.
- Product codes and lots
- Model number 6258425 - Serial numbers: 1001, 1004, 1008, 1010, 1012, 1015, 1018, 1019, 1021, 1026, and 1029. Model number 6379668 - Serial numbers: 1109, 1111, 1119, 1121, 1122, 1123, 1126, 1127, 1128, 1130, 1134, 1135, 1137, 1140, 1141, 1142, 1143, 1144, 1147, 1149, 1151, 1152, 1154, 1155, 1156, 1159, 1163, 1165, 1171, 1173, 1174, 1175, 1176, 1179, 1180, 1181, 1185, 1190, 1192, 1194, 1195, 1196, 1201, 1205, 1208, and 1209.