Z-0908-2010

Rubella IgG EIA test kit, containing 1 Rubella IgG Microplate, 1 Negative Control, 1 High Positive Control, 1...

Bio-Rad Laboratories Inc. / initiated 2008-08-20 / Terminated / root cause: manufacturing process

Bio-Rad Laboratories began a recall of Rubella IgG EIA test kit, containing 1 Rubella IgG Microplate, 1 Negative Control, 1 High Positive Control, 1 Low Positive Control, 1 Substrate, 1 Stop... on . The reason FDA records is "Elevated optical density (OD) values with the kit Negative Control". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0908-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Bio-Rad Laboratories Inc.
Product
Rubella IgG EIA test kit, containing 1 Rubella IgG Microplate, 1 Negative Control, 1 High Positive Control, 1 Low Positive Control, 1 Substrate, 1 Stop Reagent. Labeled: Distributed by: Bio-Rad Laboratories-Redmond, WA 98052 For the qualitative, semi-quantitative and quantitative detection of human IgG antibodies to rubella virus in human serum by enzyme immunoassay, to aid in the assessment to the patient's immunological response to rubella, and as a qualitative screening test to determine the immune status of individuals, including women of childbearing age.
Product code
LFX Enzyme Linked Immunoabsorbent Assay, Rubella
Reason for recall
Elevated optical density (OD) values with the kit Negative Control.
Root cause tag
manufacturing process
FDA root cause
Employee error
Action
An "Urgent Product Correction" letter dated August 26, 2008 was sent via overnight carrier to all customers. The letter described the affected product, problem, and action to be taken by customer. The customers are to review their current inventory of the product for any remaining kits of the specified lot number. The customer is to complete the form and fax to the Technical Support department if replacement kits were needed. The remaining kits will be replaced once the form is completed and faxed. If you have any questions, please contact Technical Support at 1-800-224-6723, options 2 then 4.
Quantity in commerce
82 kits
Distribution
Nationwide distribution: NE, NY, FL, GA, MA, OR, WI, MN, MI, MS, NC, CA, MD, MO, IA, DC
Product codes and lots
Lot F1108