Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0460-2009

GE Centricity PACS RA1000 Workstation (for diagnostic image analysis). GE Healthcare Integrated IT Solutions...

GE Healthcare / initiated 2008-09-25 / Terminated / root cause: software

GE Healthcare Integrated IT Solutions began a recall of GE Centricity PACS RA1000 Workstation (for diagnostic image analysis). GE Healthcare Integrated IT Solutions, Barrington, IL 60010. The Centricity" PACS... on . The reason FDA records is "Software anomalies result in patient safety issues involving patient jacket content intermittently becoming unintentionally out of synchronization with the images being displayed, and involving Default Display Protocols...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0460-2009
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
GE Healthcare
Product
GE Centricity PACS RA1000 Workstation (for diagnostic image analysis). GE Healthcare Integrated IT Solutions, Barrington, IL 60010. The Centricity" PACS Workstation is intended for use as a primary diagnostic and analysis tool for diagnostic images by trained healthcare professionals. It is also intended for use as a clinical review workstation throughout the healthcare facility. The workstation interface provides the user with a means to display, manipulate, archive, print, and export images when connected with the Centricity" PACS infrastructure.
Product code
LLZ System, Image Processing, Radiological
Reason for recall
Software anomalies result in patient safety issues involving patient jacket content intermittently becoming unintentionally out of synchronization with the images being displayed, and involving Default Display Protocols (DDPs) which are used to layout images when displaying a study. If the user does not check the Study Date Time of the exam on the image title bar, then they may interpret the curr
Root cause tag
software
FDA root cause
Software design
Action
GE Healthcare Integrated IT Solutions notified their Centricity PACS RA1000 Workstation customers via letter (Product Safety Notification) dated 9/25/08 of the software anomalies resulting in patient safety issues involving the patient jacket content intermittently becoming unintentionally out of synchronization with the images being displayed, and involving Default Display Protocols (DDPs) which are used to layout images when displaying a study. If the user does not check the Study Date Time of the exam on the image title bar, then they may interpret the current exam as a historical study and vice versa. The letter provided the users with safety instructions to follow until the software is updated. Direct any questions to the Remote Online Center, Field Engineer or the GE Customer Care Center at 800-437-1171. The accounts were requested to complete the enclosed customer reply form, supplying the workstation software or build number, and fax the completed form to 847-277-5240.
Quantity in commerce
546 units
Distribution
Worldwide Distribution including USA and countries of Australia, Belgium, Brazil, Canada, Cayman Islands, Chile, Denmark, Ecuador, France, Germany, Hong Kong, Italy, Japan, Korea, Malaysia, Malta, Mexico, New Zealand, Singapore, Spain, Switzerland, Taiwan, United Kingdom and Venezuela.
Product codes and lots
GE Centricity PACS RA1000 Workstation software versions:  1) Software version 2.1.X (from 2.1.2.1 to 2.1.5.4 inclusive for the patient jacket issue and all 2.1.x releases prior to 2.1.5.1 for the default display protocol issue), and   2) Software version 3.0.X (from 3.0.0.0 to 3.0.3.1 inclusive for the patient jacket issue and all 3.0.x releases prior to 3.0.2 for the default display protocol issue).