Platinum Plus" LT , REF 46-602, UPN M001466020, LT/0.018/260, Sterilized with ethylene oxide gas, Rx only...
Boston Scientific Corporation began a recall of Platinum Plus" LT , REF 46-602, UPN M001466020, LT/0.018/260, Sterilized with ethylene oxide gas, Rx only, Legal Manufacturer Boston Scientific Corporation... on . The reason FDA records is "Boston Scientific is initiating a voluntary recall of the Platinum Plus" Guidewire (peripheral). BSC has received reports that product may be labeled as a 260 CM long guidewire when the actual packaged device is a 180...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-0031-2009
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Boston Scientific Corporation
- Product
- Platinum Plus" LT , REF 46-602, UPN M001466020, LT/0.018/260, Sterilized with ethylene oxide gas, Rx only, Legal Manufacturer Boston Scientific Corporation, One Boston scientific Place, Natick, MA 01760-1537, Made in USA: 8600 NW 41st Street, Miami, FL 33166-6202. The Platinum Plus" Guidewire facilitates placement of a catheter during diagnostic or interventional intravascular procedures. The wire can be torqued to facilitate navigation through tortuous arteries and/or avoid unwanted side branches.
- Product code
- DQX Wire, Guide, Catheter
- Reason for recall
- Boston Scientific is initiating a voluntary recall of the Platinum Plus" Guidewire (peripheral). BSC has received reports that product may be labeled as a 260 CM long guidewire when the actual packaged device is a 180 CM long guidewire and the product may be labeled as a 180 CM long guidewire when the actual packaged device is a 260 CM long guidewire. There is no expected potential for injury to
- Root cause tag
- labeling and UDI
- FDA root cause
- Labeling mix-ups
- Action
- A "Boston Scientific Urgent Voluntary Medical Device Recall or Correction - Immediate Action Required" letter dated 8/25/08 was sent to each affected account (To The Hospital Field Action Contact). The letter includes the customer steps for recall, a complete listing of all product batches/lots being recalled, the Reply Verification Tracking Form(s), and a return shipping label. Contact Boston Scientific Corporation at 1-763-494-1700 for assistance.
- Quantity in commerce
- 35
- Distribution
- USA: CA FL GA IL IN LA MN MO NJ NV NY OH PA TN TX UT WA WI and WV OUS: France, Germany, Great Britain, Italy, and the Netherlands.
- Product codes and lots
- Lot # 11574628