Z-0218-2009

Gebauer's Fluro-Ethyl Nonflammable Topical Anesthetic Skin Refrigerant (Aerosol Can) P/N 0386-0020-20. The...

Gebauer Co / initiated 2008-09-03 / Terminated / root cause: design

Gebauer Company began a recall of Gebauer's Fluro-Ethyl Nonflammable Topical Anesthetic Skin Refrigerant (Aerosol Can) P/N 0386-0020-20. The product consists of a can, a valve and an actuator... on . The reason FDA records is "Some units of Fluro-Ethyl containing the defective valve which could malfunction and spray refrigerant out from the side of the valve in addition to or instead of spraying in the normal inverted position from the...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0218-2009
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Gebauer Co
Product
Gebauer's Fluro-Ethyl Nonflammable Topical Anesthetic Skin Refrigerant (Aerosol Can) P/N 0386-0020-20. The product consists of a can, a valve and an actuator. Gebauer's Fluro-Ethyl is a vapocoolant (skin refrigerant), topical anesthetic intended to control the pain associated with minor surgical procedures, dermabrasion and injection. It is also effective in providing temporary relief from the pain associated with minor sports injuries.
Product code
MLY Vapocoolant Device
Reason for recall
Some units of Fluro-Ethyl containing the defective valve which could malfunction and spray refrigerant out from the side of the valve in addition to or instead of spraying in the normal inverted position from the product's actuator. The product can also leak coolant from the side of the valve onto the fingers of the user.
Root cause tag
design
FDA root cause
Component design/selection
Action
A recall notification letter (Recall Notice) (dated 9/3/2008) was issued on 9/16/2008 to customers. The notice instructs customers to check their inventory for any remaining recall product and if found, return the product to the Gebauer Company. The letter also advises distributors to notify their customers and follow instructions as noted above, i.e. customers inventory their stock for recalled product and if found, return to the Gauber Company. For questions or additional information, contact Gebauer Company at 216-581-3030 (ext. 120) or 800-321-9348.
Quantity in commerce
12,717 Units
Distribution
Nationwide Distribution.
Product codes and lots
Model Number: P/N 0386-0002-20. Lot Numbers: 7010, 7011, and 7012.