ABL7XX Calibration 1 Solution S1720. P/N S1720, 944-024 Intended for in vitro testing of samples of whole...
Radiometer America Inc began a recall of ABL7XX Calibration 1 Solution S1720. P/N S1720, 944-024 Intended for in vitro testing of samples of whole blood for the parameters pH, pO2, pCO2, potassium... on . The reason FDA records is "The firm found that for this particular lot that the barcode, used for entering the characteristics of CAL1 Solution into the analyzer, does not reflect the actual values of the solution. As a consequence, the...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0457-2009
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Radiometer America, Inc
- Product
- ABL7XX Calibration 1 Solution S1720. P/N S1720, 944-024 Intended for in vitro testing of samples of whole blood for the parameters pH, pO2, pCO2, potassium, sodium, calcium, chloride, glucose, lactate, total bilirubin, and co-oximetry parameters (total hemoglobin, oxygen saturation, and the hemoglobin fractions FO2Hb, FCOHb, FMetHb, FHHb, FHbF). Also, intended for in vitro testing of samples of expired air for the parameters pO2 and pCO2.
- Product code
- JFP Electrode, Ion Specific, Calcium
- Reason for recall
- The firm found that for this particular lot that the barcode, used for entering the characteristics of CAL1 Solution into the analyzer, does not reflect the actual values of the solution. As a consequence, the calibration curves for these parameters and thereby the measured results for patient samples and quality controls will be biased.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- On, 9/2/2008 the firm sent out recall notification letters requesting their customers discard any Calibration 1 Solution (CAL1) S1720 (part number 944-024), lot WA-04, complete the attached return form and fax it to 800-736-0601. The firm will send out replacements for all affected product that the firm shows was shipped to their location. Further questions could be directed to Technical Support at 800-736-0600, option 2.
- Quantity in commerce
- 3,200/200ml bottles
- Distribution
- Worldwide Distribution --- USA including states of AZ, CA, FL, IA, ID, IL, IN, KY, LA, MD, NC, ND, NE, NY, OH, OK, PA, SD, TX, VA, WA, and WI, and country of Canada.
- Product codes and lots
- P/N S1720, 944-024; Lot: WA-04