Applied Medical, Lap Roux-EN-Y Kit, REF: K2129 Contains: (1) C0Q04, 5x100mm Threaded Kii Access System (1)...
Applied Medical Resources Corp began a recall of Applied Medical, Lap Roux-EN-Y Kit, REF: K2129 Contains: (1) C0Q04, 5x100mm Threaded Kii Access System (1) C0130, 12x100mm Optical Separator System... on . The reason FDA records is "Potential inability to insufflate through the stopcock. Under certain circumstances, a flexible elastomeric component inside the 15mm trocar can stretch into a configuration that blocks the flow of insufflation gas (C02)". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0029-2009
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Applied Medical Resources Corp
- Product
- Applied Medical, Lap Roux-EN-Y Kit, REF: K2129 Contains: (1) C0Q04, 5x100mm Threaded Kii Access System (1) C0130, 12x100mm Optical Separator System, Non-handled (2) C0Q10, 5x100mm Threaded Kii Cannula and Seal (1) C0605, 15mmx100mm Threaded Cannula System, for laparoscopic procedures.
- Product code
- GCJ Laparoscope, General & Plastic Surgery
- Reason for recall
- Potential inability to insufflate through the stopcock. Under certain circumstances, a flexible elastomeric component inside the 15mm trocar can stretch into a configuration that blocks the flow of insufflation gas (C02).
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Customers were notified via FedEx or UPS on June 25,2008 with an overnight letter. The letter advises that Applied Medical is conducting a voluntary recall of the 15mm Separator Access Systems, Models C0604, C0605, C0606 and C0607 due to the potential inability to insufflate through the stopcock. Per the letter, the firm is asking that all 15mm product listed in the letter be returned. A response form is provided.
- Quantity in commerce
- 14,113 total
- Distribution
- Nationwide and Canada distribution.
- Product codes and lots
- Product/Kit Batch Numbers: 1059636, 1060354, 1061309, 1062314