Z-0125-2011
Allura Xper FD10/10 Biplane X-ray system. Intended for use in cardiovascular and vascular x-ray imaging...
Philips North America LLC / initiated 2008-08-21 / Terminated / root cause: software
Philips Medical Systems North America Co. Phillips began a recall of Allura Xper FD10/10 Biplane X-ray system. Intended for use in cardiovascular and vascular x-ray imaging applications, including diagnostic, interventional... on . The reason FDA records is "Potential for table and stand not to move due to a software error". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0125-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Philips North America LLC
- Product
- Allura Xper FD10/10 Biplane X-ray system. Intended for use in cardiovascular and vascular x-ray imaging applications, including diagnostic, interventional procedures (such as PTCA and stent placement and atherectomies), pacemaker implantations and electrophysiology. It is compatible with specified magnetic navigation systems.
- Product code
- IZI System, X-Ray, Angiographic
- Reason for recall
- Potential for table and stand not to move due to a software error.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Philips Medical Systems North America Co. sent an URGENT - Device Correction, dated September 15, 2010, to their consignees. The letter identified the product, the problem, and the action to be taken. The consignees were informed that there was a potential geometry software lock up during a Digital Rotational Angiography (DRA) run, the stand and table movement may no longer be possible. The consignees were advised that as geometry software errors happen intermittently, no actions can be performed by the consignees to minimize the risk except for not performing DRA procedures until the firm upgrades the software. For question consignees were to contact the Philips Call Center at 800-722-9377 #5, #4, #1 and reference FCO 72200111 or FCO 72200125.
- Quantity in commerce
- 26 units
- Distribution
- Devices were distributed to hospitals and medical facilities through out the U.S.
- Product codes and lots
- Devices are identified with Site Numbers: 103281, 103283, 505903, 506119, 530653, 536715, 543411, 545396, 547823, 549368, 552969, 554802, 554970, 555293, 555299, 555834, 41443647, 41443804, 41444650, 41445448, 41445501, 41445538, 41455920, 41780160, 42057324, and 42064724.
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.
