Z-0088-2009

Envision E700 Low Airloss Therapy Surface, Made By Hill-Rom Manufacturing, Inc., Charleston, SC. The Envision...

Hill-Rom Manufacturing, Inc. / initiated 2008-06-19 / Terminated / root cause: software

Hill-Rom Manufacturing, Inc. began a recall of Envision E700 Low Airloss Therapy Surface, Made By Hill-Rom Manufacturing, Inc., Charleston, SC. The Envision E700 Low Airloss Therapy Surface helps prevent... on . The reason FDA records is "A defect in the software of the device may not allow the patient bed exam alarm or the patient movement alarm to function correctly". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0088-2009
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Hill-Rom Manufacturing, Inc.
Product
Envision E700 Low Airloss Therapy Surface, Made By Hill-Rom Manufacturing, Inc., Charleston, SC. The Envision E700 Low Airloss Therapy Surface helps prevent and treat stage III and stage IV pressure ulcers in patients who weigh between 70 lb and 400 lb and are between 4'11" and 6' 4" in height.
Product code
IOQ Bed, Flotation Therapy, Powered
Reason for recall
A defect in the software of the device may not allow the patient bed exam alarm or the patient movement alarm to function correctly.
Root cause tag
software
FDA root cause
Software change control
Action
An internal Modification Notice dated 9/1/08 was sent via email to Service Centers (because the device is owned by the manufacturer, Hill-Rom, each Service Center in possession of the device will be the consignee). Each Service Center will perform the correction on the units affected. Each device in the affected serial number range will be tested for the software defect. If the defect is found, the circuit board that contains the defective software will be replaced. The Field Action Plan is expected to be completed on July 24, 2008. Expected completion date of the Correction will be 12 months from July, 2008.
Quantity in commerce
779 units
Distribution
Worldwide Distribution --- including USA and country of Canada.
Product codes and lots
Serial numbers DB100527 thru DB101305.