Z-0102-2009
Syngo Imaging with versions VB20B, VB20D, VB20F, VB20G, VB25B, VB30A, VB30A_SP1, and VB30A_SP2
Siemens Medical Solutions USA, Inc. / initiated 2008-08-22 / Terminated / root cause: software
Siemens Medical Solutions USA, Inc began a recall of Syngo Imaging with versions VB20B, VB20D, VB20F, VB20G, VB25B, VB30A, VB30A_SP1, and VB30A_SP2 on . The reason FDA records is "Image may not be visible after merge". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0102-2009
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Product
- Syngo Imaging with versions VB20B, VB20D, VB20F, VB20G, VB25B, VB30A, VB30A_SP1, and VB30A_SP2.
- Product code
- LLZ System, Image Processing, Radiological
- Reason for recall
- Image may not be visible after merge.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- The recalling firm issued Customer Safety Advisory Notice dated 8/22/08 to inform the consignees of the potential issue and provide instruction to avoid its occurrence. To avoid potential risk of this issue, users are asked to check whether the extended HL7 messages as stated in the letter are mandatory for their RIS-PACS workflow, and disable these messages if they are not. Siemens is preparing a modification under the UI IM086/08/S for the affected versions of syngo Imaging that will resolve the potential malfunction. The field modification will be immediately distributed and installed by the firm's Service as soon as it is available.
- Quantity in commerce
- 59 units
- Distribution
- Nationwide Distribution --- including states of AL, AZ, CA, CO, CT, FL, IA, KS, LA, MA, MD, MI, MN, MO, NC, NE, NJ, NY, OH, OK, OR, PA, TN, TX, and WI.
- Product codes and lots
- Model number 10014063. Serial numbers: 1005, 1039, 1040, 1041, 1049, 1053, 1055, 1056, 1057, 1058, 1061, 1064, 1065, 1067, 1071, 1074, 1076, 1079, 1080, 1081, 1082, 1083, 1084, 1086, 1089, 1090, 1095, 1105, 1107, 1112, 1116, 1119, 1124, 1174, 1188, 1198, 1202, 1203, 1210, 1218, 1226, 1227, 1229, 1240, 1246, 1247, 1258, 1269, 1271, 1278, 1279, 1282, 1316, 1322, 1329, 1334, 1343, 1351, and 1353.