Z-0464-2009

Casmed 500cc Unifusor II with aneroid gauge and thumbwheel valve Model: 2803SGA. It is designed for use in...

Cas Medical Systems, Inc. / initiated 2008-09-20 / Terminated / root cause: manufacturing process

CAS Medical Systems, Inc. began a recall of Casmed 500cc Unifusor II with aneroid gauge and thumbwheel valve Model: 2803SGA. It is designed for use in rapid infusion,invasive pressure monitoring and with... on . The reason FDA records is "Leaks: Infusion Cuffs may exhibit air leakage at the union of the tube and bag, and not maintain pressure. For further information, please contact the firm at 203-488-6056". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0464-2009
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Cas Medical Systems, Inc.
Product
Casmed 500cc Unifusor II with aneroid gauge and thumbwheel valve Model: 2803SGA. It is designed for use in rapid infusion,invasive pressure monitoring and with auto-transfusion devices.
Product code
KZD Infusor, Pressure, For I.V. Bags
Reason for recall
Leaks: Infusion Cuffs may exhibit air leakage at the union of the tube and bag, and not maintain pressure. For further information, please contact the firm at 203-488-6056.
Root cause tag
manufacturing process
FDA root cause
Equipment maintenance
Action
CAS Medical Systems notified accounts by letter dated 9/19/08 requesting return and sub-recall be conducted to their accounts. The letter was entitled "UNIFUSOR(R) INFUSION CUFFS AND PRIVATE LABEL (CASMED-MANUFACTURED) INFUSION CUFFS *** URGENT PRODUCT CORRECTION NOTICE." For further information, please contact the firm at 203-488-6056.
Quantity in commerce
1,296 pcs
Distribution
Nationwide Foreign: Malaysia, Austria, Belgium, Czech Rep., France, Germany, Greece, Italy, Norway, Spain, Sweden, Switzerland and the UK.
Product codes and lots
Lot Numners:25380908, 41312108, and 41742108.