Z-0191-2009
Stryker Osteosynthesis Reduction Spoon; 1 single unit to a package; non-sterile; Stryker Trauma GmbH, 2432...
Howmedica Osteonics Corp / initiated 2008-08-14 / Terminated / root cause: design
Stryker Howmedica Osteonics Corp. began a recall of Stryker Osteosynthesis Reduction Spoon; 1 single unit to a package; non-sterile; Stryker Trauma GmbH, 2432 Schonkirchen, Germany. Distributed in the US by... on . The reason FDA records is "Stryker has become aware of the potential for breakage of the Reduction Spoon designed prior to 2003". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0191-2009
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Howmedica Osteonics Corp
- Product
- Stryker Osteosynthesis Reduction Spoon; 1 single unit to a package; non-sterile; Stryker Trauma GmbH, 2432 Schonkirchen, Germany. Distributed in the US by Howmedica Osteonics Corp., 325 Corporate Drive, Mahwah, NJ 007430 Reduction spoons are sold separately or as part of a kit. The Universal Rod with Reduction Spoon may be used as a fracture reduction tool to facilitate Guide Wire insertion.
- Product code
- LXH Orthopedic Manual Surgical Instrument
- Reason for recall
- Stryker has become aware of the potential for breakage of the Reduction Spoon designed prior to 2003.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Recall notification letters were sent on August 14, 2008 to all branches and hospitals via Federal Express. Questions are to be directed to 201-831-5718 or a Trauma team member listed in the letter.
- Quantity in commerce
- 1070 units
- Distribution
- Nationwide
- Product codes and lots
- Product Code 1806-0125 All product made before design change in 2003. Refer to picture provided showing differences in device.