Z-2034-2008
BI-700-00027 0-arm 1000 Imaging System Mobile X-ray System Generating 2D and 3D images of human anatomy for...
Medtronic Navigation, Inc / initiated 2008-06-30 / Terminated
Medtronic Navigation, Inc. began a recall of BI-700-00027 0-arm 1000 Imaging System Mobile X-ray System Generating 2D and 3D images of human anatomy for surgical applications on . The reason FDA records is "Failure to of affected units to comply with EER/AKR limits due to misinterpretation of the measurement requirements specified in 21 CFR 1020.32 (d) (3) (iii)". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-2034-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Medtronic Navigation, Inc
- Product
- BI-700-00027 0-arm 1000 Imaging System Mobile X-ray System Generating 2D and 3D images of human anatomy for surgical applications.
- Product code
- JAA System, X-Ray, Fluoroscopic, Image-Intensified
- Reason for recall
- Failure to of affected units to comply with EER/AKR limits due to misinterpretation of the measurement requirements specified in 21 CFR 1020.32 (d) (3) (iii)
- FDA root cause
- Radiation Control for Health and Safety Act
- Action
- Medtronic Navigation notified service engineers to retest units for compliance.
- Quantity in commerce
- 10 units
- Distribution
- CA, DC, LA, FL. IN, OH, PA, VA
- Product codes and lots
- S/N: 102, 125, 127, 129, 130, 134, 137, 147, 157, and 159.