Z-2034-2008

BI-700-00027 0-arm 1000 Imaging System Mobile X-ray System Generating 2D and 3D images of human anatomy for...

Medtronic Navigation, Inc / initiated 2008-06-30 / Terminated

Medtronic Navigation, Inc. began a recall of BI-700-00027 0-arm 1000 Imaging System Mobile X-ray System Generating 2D and 3D images of human anatomy for surgical applications on . The reason FDA records is "Failure to of affected units to comply with EER/AKR limits due to misinterpretation of the measurement requirements specified in 21 CFR 1020.32 (d) (3) (iii)". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-2034-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Medtronic Navigation, Inc
Product
BI-700-00027 0-arm 1000 Imaging System Mobile X-ray System Generating 2D and 3D images of human anatomy for surgical applications.
Product code
JAA System, X-Ray, Fluoroscopic, Image-Intensified
Reason for recall
Failure to of affected units to comply with EER/AKR limits due to misinterpretation of the measurement requirements specified in 21 CFR 1020.32 (d) (3) (iii)
FDA root cause
Radiation Control for Health and Safety Act
Action
Medtronic Navigation notified service engineers to retest units for compliance.
Quantity in commerce
10 units
Distribution
CA, DC, LA, FL. IN, OH, PA, VA
Product codes and lots
S/N: 102, 125, 127, 129, 130, 134, 137, 147, 157, and 159.