Cardiac Imaging Catheter Model #'s: ACU08267996 and ACU55790 Intended for intravascular or intracardiac...
Sterilmed Inc began a recall of Cardiac Imaging Catheter Model #'s: ACU08267996 and ACU55790 Intended for intravascular or intracardiac ultrasound imaging in order to provide visualization of... on . The reason FDA records is "SteriMed, Inc is conducting a product removal of a small group of devices individually packaged and sealed within a pouch. The removal is being conducted because the possibility exists that some of the pouches were not...". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-0551-2009
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Sterilmed Inc
- Product
- Cardiac Imaging Catheter Model #'s: ACU08267996 and ACU55790 Intended for intravascular or intracardiac ultrasound imaging in order to provide visualization of vascular anatomy, cardiac and great vessel anatomy and physiology, or other devices in the heart.
- Product code
- NLI Catheter, Angiography, Reprocessed
- Reason for recall
- SteriMed, Inc is conducting a product removal of a small group of devices individually packaged and sealed within a pouch. The removal is being conducted because the possibility exists that some of the pouches were not properly sealed on one end. A breach in packaging seal or a failure in packaging integrity has the possibility of risk to the patient in terms of transmitting organisms capable
- Root cause tag
- packaging
- FDA root cause
- Packaging process control
- Action
- Consignees were notified of the recall via verbal communication in person by on-site technician and or their local sales representative on 8/8/08. Others were contacted via fax or e-mail followed by on-site visits 8/11/08. SteriMed also provided consignees (Materials Manager/Risk Manager) with a "Urgent Product Removal" recall letter (August 14, 2008) in person informing them of the potential of receiving affected product. Included in the Product Removal Notification was a Customer Reply Form to inventory quantity of product received and quantity returned. Contact SterilMed, Inc. at 1-888-541-0078 for assistance.
- Quantity in commerce
- 58
- Distribution
- AL, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, KY, MA, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, WA, WI, and WV OUS: CANADA
- Product codes and lots
- n/a